Background: The stereolithographic-guided surgery system involves a sequence of diagnostic and therapeutic events, and errors can arise at different stages. In these systems, one of the potentially clinically relevant errors may be the mechanical errors caused by the bur-guide gap due to the presence of a rotational allowance of the drills in the tubes. Purpose: The purpose of this retrospective clinical study is to determine if it is possible to reduce the total error by limiting the tolerance among the mechanical components and to evaluate its clinical incidence. Materials and Methods: Sixty-six implants were inserted using the External Hex Safe® (Materialise Dental, Leuven, Belgium) system (Group A), and 71 implants were inserted using the same system with mechanical components modified to minimize the tolerance (Group B). Regarding only the angular deviation values, the t-test was used to determine the influence of reduced tolerance among the mechanical components on the accuracy values. Results: t-Test showed that there is a statistically significant better accuracy with the modified system (Group B). Conclusions: Limiting the error that originates from mechanical components, total error could be statistically significantly reduced. Mechanical error is one of the most important source of error using External Hex Safe stereolithographic surgical guide. © 2013 Wiley Periodicals, Inc.

The influence of the tolerance between mechanical components on the accuracy of implants inserted with a stereolithographic surgical guide: a retrospective clinical study / Cassetta, Michele; Di Mambro, Alfonso; DI GIORGIO, Gianni; Stefanelli, Luigi V.; Barbato, Ersilia. - In: CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH. - ISSN 1523-0899. - ELETTRONICO. - 17:3(2015), pp. 580-588. [10.1111/cid.12120]

The influence of the tolerance between mechanical components on the accuracy of implants inserted with a stereolithographic surgical guide: a retrospective clinical study

Michele Cassetta
Primo
;
Gianni Di Giorgio;Luigi V. Stefanelli
Penultimo
;
Ersilia Barbato
Ultimo
2015

Abstract

Background: The stereolithographic-guided surgery system involves a sequence of diagnostic and therapeutic events, and errors can arise at different stages. In these systems, one of the potentially clinically relevant errors may be the mechanical errors caused by the bur-guide gap due to the presence of a rotational allowance of the drills in the tubes. Purpose: The purpose of this retrospective clinical study is to determine if it is possible to reduce the total error by limiting the tolerance among the mechanical components and to evaluate its clinical incidence. Materials and Methods: Sixty-six implants were inserted using the External Hex Safe® (Materialise Dental, Leuven, Belgium) system (Group A), and 71 implants were inserted using the same system with mechanical components modified to minimize the tolerance (Group B). Regarding only the angular deviation values, the t-test was used to determine the influence of reduced tolerance among the mechanical components on the accuracy values. Results: t-Test showed that there is a statistically significant better accuracy with the modified system (Group B). Conclusions: Limiting the error that originates from mechanical components, total error could be statistically significantly reduced. Mechanical error is one of the most important source of error using External Hex Safe stereolithographic surgical guide. © 2013 Wiley Periodicals, Inc.
2015
computer-assisted; retrospective; stereo lithography; clinical study; implant; cad/cam technology; implantology; flapless implant surgery; accuracy
01 Pubblicazione su rivista::01a Articolo in rivista
The influence of the tolerance between mechanical components on the accuracy of implants inserted with a stereolithographic surgical guide: a retrospective clinical study / Cassetta, Michele; Di Mambro, Alfonso; DI GIORGIO, Gianni; Stefanelli, Luigi V.; Barbato, Ersilia. - In: CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH. - ISSN 1523-0899. - ELETTRONICO. - 17:3(2015), pp. 580-588. [10.1111/cid.12120]
File allegati a questo prodotto
File Dimensione Formato  
Cassetta_Influence_2015.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 718.44 kB
Formato Adobe PDF
718.44 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/521637
Citazioni
  • ???jsp.display-item.citation.pmc??? 8
  • Scopus 38
  • ???jsp.display-item.citation.isi??? 38
social impact