Background: Performing endoscopic submucosal dissection (ESD) by using standard endoscopy platforms is technically challenging because of the equipment's lack of dexterity. Objective: To explore the feasibility of using the Master and Slave Transluminal Endoscopic Robot (MASTER), a novel robotics-enhanced endosurgical system, to perform ESD. Design: ESD was performed on simulated gastric lesions in 5 Erlangen porcine stomach models (ex vivo) and 5 live pigs (in vivo). Performance of ESD by using the MASTER was compared with that using the insulation-tipped (IT) diathermic knife. Setting: SMART Laboratory, Advance Surgical Training Centre, National University Hospital, Singapore. Subjects: Five Erlangen porcine stomach models and 5 pigs, 5 to 7 months old, each weighing about 35 kg. Interventions: ESD. Main Outcome Measurements: Lesion resection time, grasper and hook efficacy grade, completeness of resection, and presence of procedure-related perforation. Results: In the Erlangen stomach models, 15 simulated lesions from the cardia, antrum, and body were removed en bloc (mean dimension, 37.4 X 26.5 mm) by electrocautery excision using the MASTER. The mean ESD time was 23.9 minutes (range 7-48 minutes). There was no difference in the dissection times of lesions at different locations (P = .449). In the live pigs, the MASTER took a mean of 16.2 minutes (range 3-29 minutes) to complete the ESD of 5 gastric lesions, whereas the IT diathermic knife took 18.6 minutes (range 9-34 minutes). There was no significant difference in the times taken (P = .708). All lesions were excised en bloc; the mean dimensions of lesions resected by the MASTER and the IT diathermic knife were 37.2 X 30.1 mm and 32.78 X 25.6 mm, respectively. The MASTER exhibited good grasping and cutting efficiency throughout. Surgical maneuvers were achieved with ease and precision. There was no incidence of excessive bleeding or stomach wall perforation. Limitations: Exploratory study with limited sample size. Conclusions: Performing ESD by using the MASTER is feasible. (Gastrointest Endosc 2010;72:593-9)

Endoscopic submucosal dissection of gastric lesions by using a Master and Slave Transluminal Endoscopic Robot (MASTER) / Khek Yu, Ho; Soo Jay, Phee; Asim, Shabbir; Soon Chiang, Low; Van An, Huynh; Andy Prima, Kencana; Kai, Yang; Lomanto, Davide; B. Y., Jimmy; Y. Y., Jennie; S. C., Sydney. - In: GASTROINTESTINAL ENDOSCOPY. - ISSN 0016-5107. - 72:3(2010), pp. 593-599. [10.1016/j.gie.2010.04.009]

Endoscopic submucosal dissection of gastric lesions by using a Master and Slave Transluminal Endoscopic Robot (MASTER)

LOMANTO, Davide;
2010

Abstract

Background: Performing endoscopic submucosal dissection (ESD) by using standard endoscopy platforms is technically challenging because of the equipment's lack of dexterity. Objective: To explore the feasibility of using the Master and Slave Transluminal Endoscopic Robot (MASTER), a novel robotics-enhanced endosurgical system, to perform ESD. Design: ESD was performed on simulated gastric lesions in 5 Erlangen porcine stomach models (ex vivo) and 5 live pigs (in vivo). Performance of ESD by using the MASTER was compared with that using the insulation-tipped (IT) diathermic knife. Setting: SMART Laboratory, Advance Surgical Training Centre, National University Hospital, Singapore. Subjects: Five Erlangen porcine stomach models and 5 pigs, 5 to 7 months old, each weighing about 35 kg. Interventions: ESD. Main Outcome Measurements: Lesion resection time, grasper and hook efficacy grade, completeness of resection, and presence of procedure-related perforation. Results: In the Erlangen stomach models, 15 simulated lesions from the cardia, antrum, and body were removed en bloc (mean dimension, 37.4 X 26.5 mm) by electrocautery excision using the MASTER. The mean ESD time was 23.9 minutes (range 7-48 minutes). There was no difference in the dissection times of lesions at different locations (P = .449). In the live pigs, the MASTER took a mean of 16.2 minutes (range 3-29 minutes) to complete the ESD of 5 gastric lesions, whereas the IT diathermic knife took 18.6 minutes (range 9-34 minutes). There was no significant difference in the times taken (P = .708). All lesions were excised en bloc; the mean dimensions of lesions resected by the MASTER and the IT diathermic knife were 37.2 X 30.1 mm and 32.78 X 25.6 mm, respectively. The MASTER exhibited good grasping and cutting efficiency throughout. Surgical maneuvers were achieved with ease and precision. There was no incidence of excessive bleeding or stomach wall perforation. Limitations: Exploratory study with limited sample size. Conclusions: Performing ESD by using the MASTER is feasible. (Gastrointest Endosc 2010;72:593-9)
2010
animal; animals; computer-assisted; disease models; dissection; electrocoagulation; equipment design; feasibility studies; gastric mucosa; gastroscopes; instrumentation; pathology/surgery; robotics; stomach neoplasms; surgery; surgical instruments; swine
01 Pubblicazione su rivista::01a Articolo in rivista
Endoscopic submucosal dissection of gastric lesions by using a Master and Slave Transluminal Endoscopic Robot (MASTER) / Khek Yu, Ho; Soo Jay, Phee; Asim, Shabbir; Soon Chiang, Low; Van An, Huynh; Andy Prima, Kencana; Kai, Yang; Lomanto, Davide; B. Y., Jimmy; Y. Y., Jennie; S. C., Sydney. - In: GASTROINTESTINAL ENDOSCOPY. - ISSN 0016-5107. - 72:3(2010), pp. 593-599. [10.1016/j.gie.2010.04.009]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/400024
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 19
  • Scopus 83
  • ???jsp.display-item.citation.isi??? 71
social impact