An improved method is reported for determination of Polycyclic Aromatic Hydrocarbons (PAHs) in atmospheric particulate matter by HPLC—FLD. The sampling step (air volume collected during each sampling period varies in the range 10≑13 m3) is carried out by means of a mediumflow pumping system (15 L min—1) on a glass fiber filter (47 mm diameter) placed as collecting substrate in the samplingcassette. After sampling, the filter is extracted with 3 ml of acetonitrile in an ultrasonic bath for 30 minutes. As for extraction of PAHs from loaded filters a new criterion here is proposed to evaluate the recovery efficiency of PAHs from the sample, instead of the usual spiking method of standard solution. The extract was then reduced to 100 μL and analysed by HPLC—FLD on line spectra system. The method is rapid (about one hour for extraction and analysis), reproducible and enables to measure with good accuracy the atmospheric concentration of benzo(b) fluoranthene (BbF), benzo(k)fluoranthene (BkF), benzo(a)pyrene (BaP), benzo(ghi) perylene (BghiP), carcinogenic compounds always present in the urban airborne particulate matter. So it is useful for routine pollution studies and suitable to substitute the official method used now. Monthly average air concentrations, for the four PAHs above mentioned, measured in Rome from July 2001 to June 2002, are reported
Improved analysis of polycyclic aromatic hydrocarbons in atmospheric particulate matter by HPLC-fluorescence / Bacaloni, Alessandro; Cafaro, Claudia; de Giorgi, Loretta; Ruocco, Roberta; Zoccolillo, Lelio. - In: ANNALI DI CHIMICA. - ISSN 0003-4592. - STAMPA. - 94:9-10(2004), pp. 751-759. [10.1002/adic.200490093]
Improved analysis of polycyclic aromatic hydrocarbons in atmospheric particulate matter by HPLC-fluorescence
Bacaloni, Alessandro;Cafaro, Claudia;RUOCCO, Roberta;Zoccolillo, Lelio
2004
Abstract
An improved method is reported for determination of Polycyclic Aromatic Hydrocarbons (PAHs) in atmospheric particulate matter by HPLC—FLD. The sampling step (air volume collected during each sampling period varies in the range 10≑13 m3) is carried out by means of a mediumflow pumping system (15 L min—1) on a glass fiber filter (47 mm diameter) placed as collecting substrate in the samplingcassette. After sampling, the filter is extracted with 3 ml of acetonitrile in an ultrasonic bath for 30 minutes. As for extraction of PAHs from loaded filters a new criterion here is proposed to evaluate the recovery efficiency of PAHs from the sample, instead of the usual spiking method of standard solution. The extract was then reduced to 100 μL and analysed by HPLC—FLD on line spectra system. The method is rapid (about one hour for extraction and analysis), reproducible and enables to measure with good accuracy the atmospheric concentration of benzo(b) fluoranthene (BbF), benzo(k)fluoranthene (BkF), benzo(a)pyrene (BaP), benzo(ghi) perylene (BghiP), carcinogenic compounds always present in the urban airborne particulate matter. So it is useful for routine pollution studies and suitable to substitute the official method used now. Monthly average air concentrations, for the four PAHs above mentioned, measured in Rome from July 2001 to June 2002, are reportedI documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.