This paper deals with the electric and magnetic field (EMF) safety aspects of a wireless power transfer (WPT) system based on magnetic resonant coupling between two coils. The primary coil is assumed to be on-body, while the secondary coil is assumed to be inside the human body and connected to a battery recharge system of an active implantable medical device such as a pacemaker. This study allows us to identify a good preliminary solution of the WPT coil configuration, compensation capacitor topology, and operational frequency. Demonstrative WPT systems operating at two different frequencies are proposed in order to verify the WPT performances. The EMF safety has been finally assessed by numerical dosimetry studies using anatomically realistic human body models revealing no particular concerns about this application.

Wireless power transfer charging system for AIMDs and pacemakers / Campi, Tommaso; Cruciani, Silvano; Palandrani, Federica; DE SANTIS, Valerio; Hirata, Akimasa; Feliziani, Mauro. - In: IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES. - ISSN 0018-9480. - 64:2(2016), pp. 633-642. [10.1109/TMTT.2015.2511011]

Wireless power transfer charging system for AIMDs and pacemakers

Campi, Tommaso;DE SANTIS, VALERIO;FELIZIANI, MAURO
2016

Abstract

This paper deals with the electric and magnetic field (EMF) safety aspects of a wireless power transfer (WPT) system based on magnetic resonant coupling between two coils. The primary coil is assumed to be on-body, while the secondary coil is assumed to be inside the human body and connected to a battery recharge system of an active implantable medical device such as a pacemaker. This study allows us to identify a good preliminary solution of the WPT coil configuration, compensation capacitor topology, and operational frequency. Demonstrative WPT systems operating at two different frequencies are proposed in order to verify the WPT performances. The EMF safety has been finally assessed by numerical dosimetry studies using anatomically realistic human body models revealing no particular concerns about this application.
2016
Electric and magnetic field (EMF) safety; Magnetic resonant coupling; Numerical dosimetry; Pacemaker; Terms-Active implantable medical device (AIMD); Wireless power transfer (WPT); Electrical and Electronic Engineering; Condensed Matter Physics; Radiation
01 Pubblicazione su rivista::01a Articolo in rivista
Wireless power transfer charging system for AIMDs and pacemakers / Campi, Tommaso; Cruciani, Silvano; Palandrani, Federica; DE SANTIS, Valerio; Hirata, Akimasa; Feliziani, Mauro. - In: IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES. - ISSN 0018-9480. - 64:2(2016), pp. 633-642. [10.1109/TMTT.2015.2511011]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1707880
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 204
  • ???jsp.display-item.citation.isi??? 167
social impact