Catheter dislocation and fracture with migration of central venous lines have been reported in the International literature. Catheter fracture with consequent migration has been observed in 0.5-3.0% and may either be consequent to catheter removal or it can occur spontane-ously. Our case report concerns the migration of a Hickman catheter connected to a venous port to the right atrium in a 61-year old patient. A literature up-to-date has been performed to assess the risk of port-a-cath positioning. The position of catheter tip is considered critical for the risk of migration, that is greater as higher the tip localization respect to the carina. The aim of our study is to underline the critical role of X-ray to visualize the exact location of the catheter tip, regard-less of the approach used for catheter positioning.

Fracture and migration in right atrium of a permanent venous central access system in a elderly patient: case report and literature review / Gazzanelli, S; Miccini, M; Sapienza, P; Cavallaro, G; Biacchi, D; Crocetti, D; Fiori, E; Sammartino, P; Guerra, C; Ranieri, M. - In: LA CLINICA TERAPEUTICA. - ISSN 1972-6007. - 173:3(2022), pp. 207-213. [10.7417/CT.2022.2419]

Fracture and migration in right atrium of a permanent venous central access system in a elderly patient: case report and literature review

S Gazzanelli
;
M Miccini;P Sapienza;G Cavallaro;D Biacchi;D Crocetti;E Fiori;P Sammartino;C Guerra;M Ranieri
2022

Abstract

Catheter dislocation and fracture with migration of central venous lines have been reported in the International literature. Catheter fracture with consequent migration has been observed in 0.5-3.0% and may either be consequent to catheter removal or it can occur spontane-ously. Our case report concerns the migration of a Hickman catheter connected to a venous port to the right atrium in a 61-year old patient. A literature up-to-date has been performed to assess the risk of port-a-cath positioning. The position of catheter tip is considered critical for the risk of migration, that is greater as higher the tip localization respect to the carina. The aim of our study is to underline the critical role of X-ray to visualize the exact location of the catheter tip, regard-less of the approach used for catheter positioning.
2022
port-a-cath dislocation; port-a-cath complications; port-a-cath migration.
01 Pubblicazione su rivista::01i Case report
Fracture and migration in right atrium of a permanent venous central access system in a elderly patient: case report and literature review / Gazzanelli, S; Miccini, M; Sapienza, P; Cavallaro, G; Biacchi, D; Crocetti, D; Fiori, E; Sammartino, P; Guerra, C; Ranieri, M. - In: LA CLINICA TERAPEUTICA. - ISSN 1972-6007. - 173:3(2022), pp. 207-213. [10.7417/CT.2022.2419]
File allegati a questo prodotto
File Dimensione Formato  
Gazzanelli_Fracture-and-migration_2022.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 166.92 kB
Formato Adobe PDF
166.92 kB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1680216
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 0
  • ???jsp.display-item.citation.isi??? ND
social impact