Background: Alto is the latest generation of the Ovation stent-graft platform for endovascular aneurysm repair (EVAR). Its ultra-low profile and its proximal sealing zone close to the lowest renal artery (≥7 mm) increase standard EVAR eligibility. We report early clinical and technical outcomes with the Alto stent-graft in our University Hospital Center after CE Mark approval in August 2020. Methods: Seven patients (all male, mean age 76.1 ± 6.2 years) underwent EVAR with Ovation Alto stent-graft between June 2021 and February 2022. All the EVAR procedures were performed by a team of vascular surgeons experienced on EVAR with previous generation of Ovation platform. Follow-up consisted of duplex ultrasound examination at 1, 3, and 6 months and of a 1-month control computed tomography angiography (CTA). Patients treated gave consent to participate in this case series and publication. A descriptive analysis of variables was performed. SPSS (version 25) and Excel were used for statistical analysis. Results: Most of the patients had a fusiform abdominal aortic aneurysm (n = 5; 71.4%). The median maximal transversal aortic diameter was 5.06 cm (range, 3.98-6.99). Because of hostile aortic neck anatomy, on-label EVAR was considered feasible only with Ovation Alto stent-graft. Narrow iliac arteries (<6 mm) were also present in 2 cases. All procedures were performed according to the instruction for the use of the device. Technical success was achieved in all cases. No type IA/IB/III endoleak occurred at completion angiography. No distal migration (>10 mm) but 2 distal displacements (≥2 mm) were observed at control CTA. During follow-up, duplex ultrasound and CTA showed no type I/III endoleak, no stent-graft migration (>10 mm), and no proximal aortic neck variations (P = not statistically significant). Three patients (42.8%) are under strict surveillance because of low-flow type II endoleak not associated with sac variations. Conclusions: Our early experience shows promising technical and clinical success with Alto stent-graft. The proximal relocation of the proximal sealing rings and the ultra-low profile delivery system allow on-label EVAR in a wider range of aortic anatomies. Notwithstanding, further studies, meta-analysis, and prospective registries are mandatory to evaluate mid- and long-term efficacy and safety of this latest Ovation platform.

Early Experience with Ovation Alto Stent-Graft / Cuozzo, S.; Martinelli, O.; Brizzi, V.; Miceli, F.; Flora, F.; Sbarigia, E.; Gattuso, R.. - In: ANNALS OF VASCULAR SURGERY. - ISSN 1615-5947. - 88:(2023), pp. 346-353. [10.1016/j.avsg.2022.07.030]

Early Experience with Ovation Alto Stent-Graft

Cuozzo S.
Investigation
;
Martinelli O.;Miceli F.
Methodology
;
Flora F.
Software
;
Sbarigia E.
Conceptualization
;
Gattuso R.
Validation
2023

Abstract

Background: Alto is the latest generation of the Ovation stent-graft platform for endovascular aneurysm repair (EVAR). Its ultra-low profile and its proximal sealing zone close to the lowest renal artery (≥7 mm) increase standard EVAR eligibility. We report early clinical and technical outcomes with the Alto stent-graft in our University Hospital Center after CE Mark approval in August 2020. Methods: Seven patients (all male, mean age 76.1 ± 6.2 years) underwent EVAR with Ovation Alto stent-graft between June 2021 and February 2022. All the EVAR procedures were performed by a team of vascular surgeons experienced on EVAR with previous generation of Ovation platform. Follow-up consisted of duplex ultrasound examination at 1, 3, and 6 months and of a 1-month control computed tomography angiography (CTA). Patients treated gave consent to participate in this case series and publication. A descriptive analysis of variables was performed. SPSS (version 25) and Excel were used for statistical analysis. Results: Most of the patients had a fusiform abdominal aortic aneurysm (n = 5; 71.4%). The median maximal transversal aortic diameter was 5.06 cm (range, 3.98-6.99). Because of hostile aortic neck anatomy, on-label EVAR was considered feasible only with Ovation Alto stent-graft. Narrow iliac arteries (<6 mm) were also present in 2 cases. All procedures were performed according to the instruction for the use of the device. Technical success was achieved in all cases. No type IA/IB/III endoleak occurred at completion angiography. No distal migration (>10 mm) but 2 distal displacements (≥2 mm) were observed at control CTA. During follow-up, duplex ultrasound and CTA showed no type I/III endoleak, no stent-graft migration (>10 mm), and no proximal aortic neck variations (P = not statistically significant). Three patients (42.8%) are under strict surveillance because of low-flow type II endoleak not associated with sac variations. Conclusions: Our early experience shows promising technical and clinical success with Alto stent-graft. The proximal relocation of the proximal sealing rings and the ultra-low profile delivery system allow on-label EVAR in a wider range of aortic anatomies. Notwithstanding, further studies, meta-analysis, and prospective registries are mandatory to evaluate mid- and long-term efficacy and safety of this latest Ovation platform.
2023
Alto stent-graftAAATASC DPioneer plus intravascular re-entry IVUS-guided catheter
01 Pubblicazione su rivista::01i Case report
Early Experience with Ovation Alto Stent-Graft / Cuozzo, S.; Martinelli, O.; Brizzi, V.; Miceli, F.; Flora, F.; Sbarigia, E.; Gattuso, R.. - In: ANNALS OF VASCULAR SURGERY. - ISSN 1615-5947. - 88:(2023), pp. 346-353. [10.1016/j.avsg.2022.07.030]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1667441
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 2
social impact