To enumerate bacteria adherent to medical devices, Vortex-Sonication-Vortex Method (VSVM) and BioTimer Assay (BTA) have been applied. VSVM counts detached microorganisms whereas BTA enumerates adherent microorganisms through microbial metabolism. However, the limitation of VSVM consists in incomplete detachment of adherent microorganisms while BTA is unable to identify microbial genera and species. Herein, the combined use of VSVM and BTA for the diagnosis and enumeration of adherent microorganisms causing implant-associated-infections (IAIs) is reported. Over 2016-2018, 46 patients with IAIs were enrolled and their 82 explanted devices were submitted firstly to VSVM and then to BTA. VSVM plus BTA detected microorganisms in 39/46 patients (84.7%) compared with 32/46 (69.5%) and 31/46 (67.3%) by VSVM and BTA alone, respectively. Likely, combined methods led to microorganism detection in 54/82 devices (65.9%) compared with each method alone [43/82 (52.4%), 44/82 (53.6%) for VSVM and BTA, respectively]. The combination of both methods (concordance 75.6%) raised the sensitivity of microbial analysis in IAIs compared with either VSVM or BTA alone, thus representing a simple and accurate way for the identification and enumeration of microorganisms adherent on devices. Moreover, BTA reagent applied in a new apparatus allowed also the enumeration of the microorganisms adherent on different segments of cardiac electrodes, thus contributing to define IAIs pathogenesis.

BioTimer assay as complementary method to vortex-sonication-vortex technique for the microbiological diagnosis of implant associated infections / Rosa, Luigi; Lepanto, Maria Stefania; Cutone, Antimo; Berlutti, Francesca; De Angelis, Massimiliano; Vullo, Vincenzo; Mastroianni, Claudio Maria; Valenti, Piera; Oliva, Alessandra. - In: SCIENTIFIC REPORTS. - ISSN 2045-2322. - 9:1(2019), pp. 1-10. [10.1038/s41598-019-44045-1]

BioTimer assay as complementary method to vortex-sonication-vortex technique for the microbiological diagnosis of implant associated infections

Rosa, Luigi;Lepanto, Maria Stefania;Cutone, Antimo;Berlutti, Francesca;De Angelis, Massimiliano;Vullo, Vincenzo;Mastroianni, Claudio Maria
Membro del Collaboration Group
;
Valenti, Piera
Conceptualization
;
Oliva, Alessandra
2019

Abstract

To enumerate bacteria adherent to medical devices, Vortex-Sonication-Vortex Method (VSVM) and BioTimer Assay (BTA) have been applied. VSVM counts detached microorganisms whereas BTA enumerates adherent microorganisms through microbial metabolism. However, the limitation of VSVM consists in incomplete detachment of adherent microorganisms while BTA is unable to identify microbial genera and species. Herein, the combined use of VSVM and BTA for the diagnosis and enumeration of adherent microorganisms causing implant-associated-infections (IAIs) is reported. Over 2016-2018, 46 patients with IAIs were enrolled and their 82 explanted devices were submitted firstly to VSVM and then to BTA. VSVM plus BTA detected microorganisms in 39/46 patients (84.7%) compared with 32/46 (69.5%) and 31/46 (67.3%) by VSVM and BTA alone, respectively. Likely, combined methods led to microorganism detection in 54/82 devices (65.9%) compared with each method alone [43/82 (52.4%), 44/82 (53.6%) for VSVM and BTA, respectively]. The combination of both methods (concordance 75.6%) raised the sensitivity of microbial analysis in IAIs compared with either VSVM or BTA alone, thus representing a simple and accurate way for the identification and enumeration of microorganisms adherent on devices. Moreover, BTA reagent applied in a new apparatus allowed also the enumeration of the microorganisms adherent on different segments of cardiac electrodes, thus contributing to define IAIs pathogenesis.
2019
Infections; diagnosis; biotimer assay
01 Pubblicazione su rivista::01a Articolo in rivista
BioTimer assay as complementary method to vortex-sonication-vortex technique for the microbiological diagnosis of implant associated infections / Rosa, Luigi; Lepanto, Maria Stefania; Cutone, Antimo; Berlutti, Francesca; De Angelis, Massimiliano; Vullo, Vincenzo; Mastroianni, Claudio Maria; Valenti, Piera; Oliva, Alessandra. - In: SCIENTIFIC REPORTS. - ISSN 2045-2322. - 9:1(2019), pp. 1-10. [10.1038/s41598-019-44045-1]
File allegati a questo prodotto
File Dimensione Formato  
Rosa_BioTimer-assay_2019.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 1.22 MB
Formato Adobe PDF
1.22 MB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1274280
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 6
social impact