OBJECTIVES: The aim of this study was to evaluate the safety and efficacy of the artificial urinary sphincter ZSI 375 device in male patients with stress urinary incontinence (SUI) at 10 centres. METHODS: This was a retrospective, nonrandomised, multicentre study. From January 2012 to December 2014, and follow-up to February 2016, men with SUI were treated with the ZSI 375 device. The primary outcome was continence. The secondary outcome was complications rate. RESULTS: A total of 147 patients with moderate-to-severe stress incontinence and mean age of 70.67 (26-85) years underwent implantation of the ZSI 375 device at 10 centres.The most common indication for placement was incontinence after radical prostatectomy (121 patients: 82.31%). The ZSI 375 device was considered successful in 86.40% of cases after 12 months follow-up, 83.52% after 24 months follow-up and 78.26% after 36 months follow-up. The infection rate was 6.12%, urethral erosion was 12.92% and mechanical failure was limited to 3.40% leading to ZSI 375 improvement. CONCLUSIONS: This study confirmed safety and efficacy of ZSI 375 device to treat moderate-to-severe SUI in men.

Multicentre experience with ZSI 375 artificial urinary sphincter for the treatment of stress urinary incontinence in men / Ostrowski, I; Blewniewski, M; Neugart, F; von Heyden, B; Selvaggio, O; Iori, Francesco; Foley, S; Fernández Arjona, M; Carvajal Obando, A; Pottek, T. 1. 0.. - In: UROLOGIA. - ISSN 0391-5603. - STAMPA. - (2017), pp. 1-5. [10.5301/uj.5000246]

Multicentre experience with ZSI 375 artificial urinary sphincter for the treatment of stress urinary incontinence in men

IORI, Francesco;
2017

Abstract

OBJECTIVES: The aim of this study was to evaluate the safety and efficacy of the artificial urinary sphincter ZSI 375 device in male patients with stress urinary incontinence (SUI) at 10 centres. METHODS: This was a retrospective, nonrandomised, multicentre study. From January 2012 to December 2014, and follow-up to February 2016, men with SUI were treated with the ZSI 375 device. The primary outcome was continence. The secondary outcome was complications rate. RESULTS: A total of 147 patients with moderate-to-severe stress incontinence and mean age of 70.67 (26-85) years underwent implantation of the ZSI 375 device at 10 centres.The most common indication for placement was incontinence after radical prostatectomy (121 patients: 82.31%). The ZSI 375 device was considered successful in 86.40% of cases after 12 months follow-up, 83.52% after 24 months follow-up and 78.26% after 36 months follow-up. The infection rate was 6.12%, urethral erosion was 12.92% and mechanical failure was limited to 3.40% leading to ZSI 375 improvement. CONCLUSIONS: This study confirmed safety and efficacy of ZSI 375 device to treat moderate-to-severe SUI in men.
2017
artificial urinary sphincter; male urinary stress incontinence; zsi 375
01 Pubblicazione su rivista::01a Articolo in rivista
Multicentre experience with ZSI 375 artificial urinary sphincter for the treatment of stress urinary incontinence in men / Ostrowski, I; Blewniewski, M; Neugart, F; von Heyden, B; Selvaggio, O; Iori, Francesco; Foley, S; Fernández Arjona, M; Carvajal Obando, A; Pottek, T. 1. 0.. - In: UROLOGIA. - ISSN 0391-5603. - STAMPA. - (2017), pp. 1-5. [10.5301/uj.5000246]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/986577
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 8
  • Scopus ND
  • ???jsp.display-item.citation.isi??? 17
social impact