Objectives: To determine the optimal window setting for displaying virtual monoenergetic reconstructions of third generation dual-source, dual-energy CT (DECT) angiography of the abdomen. Methods: Forty-five patients were evaluated with DECT angiography (90/150 kV, 180/90 ref. mAs). Three datasets were reconstructed: standard linear blending (M_0.6), 70 keV traditional virtual monoenergetic (M70), and 40 keV advanced noise-optimized virtual monoenergetic (M40+). The best window setting (width and level, W/L) was assessed by two blinded observers and was correlated with aortic attenuation to obtain the Optimized W/L setting (O-W/L). Subjective image quality was assessed, and vessel diameters were measured to determine any possible influences between different W/L settings. Repeated measures of variance were used to evaluate comparison of W/L values, image quality, and vessel sizing between M_0.6, M70, and M40+. Results: The Best W/L (B-W/L) for M70 and M40+ was 880/280 and 1410/450, respectively. Results from regression analysis inferred an O-W/L of 850/270 for M70 and 1350/430 for M40+. Significant differences for W and L were found between the Best and the Optimized W/L for M40+, and between M70 and M40+ for both the Best and Optimized W/L. No significant differences for vessel measurements were found using the O-W/L for M40+ compared to the standard M_0.6 (p ≥ 0.16), and significant differences were observed when using the B-W/L with M40+ compared to M_0.6 (p ≤ 0.04). Conclusion: In order to optimize virtual monoenergetic imaging with both traditional M70 and advanced M40+, adjusting the W/L settings is necessary. Our results suggest a W/L setting of 850/270 for M70 and 1350/430 for M40+.

Optimization of window settings for standard and advanced virtual monoenergetic imaging in abdominal dual-energy CT angiography / Caruso, Damiano; Parinella, Ashley H.; Schoepf, U. Joseph; Stroebel, Maxwell H.; Mangold, Stefanie; Wichmann, Julian L.; Varga Szemes, Akos; Ball, B. Devon; DE SANTIS, Domenico; Laghi, Andrea; DE CECCO, CARLO NICOLA. - In: ABDOMINAL RADIOLOGY. - ISSN 2366-004X. - ELETTRONICO. - 42:3(2017), pp. 772-780. [10.1007/s00261-016-0963-9]

Optimization of window settings for standard and advanced virtual monoenergetic imaging in abdominal dual-energy CT angiography

CARUSO, DAMIANO;DE SANTIS, DOMENICO;LAGHI, ANDREA;DE CECCO, CARLO NICOLA
2017

Abstract

Objectives: To determine the optimal window setting for displaying virtual monoenergetic reconstructions of third generation dual-source, dual-energy CT (DECT) angiography of the abdomen. Methods: Forty-five patients were evaluated with DECT angiography (90/150 kV, 180/90 ref. mAs). Three datasets were reconstructed: standard linear blending (M_0.6), 70 keV traditional virtual monoenergetic (M70), and 40 keV advanced noise-optimized virtual monoenergetic (M40+). The best window setting (width and level, W/L) was assessed by two blinded observers and was correlated with aortic attenuation to obtain the Optimized W/L setting (O-W/L). Subjective image quality was assessed, and vessel diameters were measured to determine any possible influences between different W/L settings. Repeated measures of variance were used to evaluate comparison of W/L values, image quality, and vessel sizing between M_0.6, M70, and M40+. Results: The Best W/L (B-W/L) for M70 and M40+ was 880/280 and 1410/450, respectively. Results from regression analysis inferred an O-W/L of 850/270 for M70 and 1350/430 for M40+. Significant differences for W and L were found between the Best and the Optimized W/L for M40+, and between M70 and M40+ for both the Best and Optimized W/L. No significant differences for vessel measurements were found using the O-W/L for M40+ compared to the standard M_0.6 (p ≥ 0.16), and significant differences were observed when using the B-W/L with M40+ compared to M_0.6 (p ≤ 0.04). Conclusion: In order to optimize virtual monoenergetic imaging with both traditional M70 and advanced M40+, adjusting the W/L settings is necessary. Our results suggest a W/L setting of 850/270 for M70 and 1350/430 for M40+.
2017
ct angiography; dual-energy ct; virtual monoenergetic; w/l; window settings; radiology, nuclear medicine and imaging; gastroenterology; urology; radiological and ultrasound technology
01 Pubblicazione su rivista::01a Articolo in rivista
Optimization of window settings for standard and advanced virtual monoenergetic imaging in abdominal dual-energy CT angiography / Caruso, Damiano; Parinella, Ashley H.; Schoepf, U. Joseph; Stroebel, Maxwell H.; Mangold, Stefanie; Wichmann, Julian L.; Varga Szemes, Akos; Ball, B. Devon; DE SANTIS, Domenico; Laghi, Andrea; DE CECCO, CARLO NICOLA. - In: ABDOMINAL RADIOLOGY. - ISSN 2366-004X. - ELETTRONICO. - 42:3(2017), pp. 772-780. [10.1007/s00261-016-0963-9]
File allegati a questo prodotto
File Dimensione Formato  
Caruso_Optimization_2017.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 1.23 MB
Formato Adobe PDF
1.23 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/961742
Citazioni
  • ???jsp.display-item.citation.pmc??? 5
  • Scopus 29
  • ???jsp.display-item.citation.isi??? 27
social impact