Stripline Beam Position Monitor (BPM) will be installed in the Compton Gamma Source in construction at the ELI Nuclear Physics facility in Romania. A test bench for the calibration of BPM has been built to characterize the device with stretched wire measurement in order to get the BPM response map. A full S-parameters characterization is performed as well to measure the electrical offset with the “Lambertson method”. This paper discusses the extensive simulations performed with full 3D electromagnetic CAD codes of the above measurements to investigate measurement accuracy, possible measurement artefacts and beam position reconstruction.

On the calibration measurement of stripline beam position monitor for the ELI-NP facility / DE ARCANGELIS, Daniele; Cardelli, Fabio; Mostacci, Andrea; Palumbo, Luigi. - ELETTRONICO. - (2016), pp. 411-413. (Intervento presentato al convegno 7th International Particle Accelerator Conference, IPAC 2016 tenutosi a BEXCO Convention Center, kor nel 2016) [10.18429/JACoW-IPAC2016-MOPMW007].

On the calibration measurement of stripline beam position monitor for the ELI-NP facility

DE ARCANGELIS, DANIELE;CARDELLI, FABIO;MOSTACCI, Andrea;PALUMBO, Luigi
2016

Abstract

Stripline Beam Position Monitor (BPM) will be installed in the Compton Gamma Source in construction at the ELI Nuclear Physics facility in Romania. A test bench for the calibration of BPM has been built to characterize the device with stretched wire measurement in order to get the BPM response map. A full S-parameters characterization is performed as well to measure the electrical offset with the “Lambertson method”. This paper discusses the extensive simulations performed with full 3D electromagnetic CAD codes of the above measurements to investigate measurement accuracy, possible measurement artefacts and beam position reconstruction.
2016
7th International Particle Accelerator Conference, IPAC 2016
Nuclear and High Energy Physics; Atomic and Molecular Physics, and Optics
04 Pubblicazione in atti di convegno::04b Atto di convegno in volume
On the calibration measurement of stripline beam position monitor for the ELI-NP facility / DE ARCANGELIS, Daniele; Cardelli, Fabio; Mostacci, Andrea; Palumbo, Luigi. - ELETTRONICO. - (2016), pp. 411-413. (Intervento presentato al convegno 7th International Particle Accelerator Conference, IPAC 2016 tenutosi a BEXCO Convention Center, kor nel 2016) [10.18429/JACoW-IPAC2016-MOPMW007].
File allegati a questo prodotto
File Dimensione Formato  
De-Arcangelis_Calibration_2016.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 985.54 kB
Formato Adobe PDF
985.54 kB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/961096
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? ND
social impact