Background: The attainment of a good primary stability is a necessary condition to ensure the success of osseointegration in implantology. In type IV cancellous bone, however, it is possible that a reduced primary stability can lead to an increased rate of failure. The aim of this study was therefore to determine, with the help of the resonance frequency (Osstell mentor), which technique of implant site preparation (piezo surgery, conventional, under-preparation, bone compaction, osteodistraction) and macro-geometry is able to improve implant stability in type IV cancellous bone. Material and Methods: 10 pig ribs were prepared with a surgical pre-drilled guide, calibrated for a correct implant positioning. On each rib, 5 implant sites (one for each technique) were prepared. Successively, 50 conical implants (Tekka Global D) were inserted and measured with the resonance frequency to evaluate the primary stability. Data collected were analyzed by analysis of variance (ANOVA) to test whether the Implant Stability Quotient (ISQ) values of the five techniques were significantly different. Results: The results showed that no significant differences among the ISQ values of the five techniques used were found. Also, no significant differences in the macro-geometry of the two types of compared implants were observed. However, the macro-geometry of Tekka implants, characterized by a double condensing thread, seems to provide greater ISQ values than those of single thread implants when using the same technique. Conclusions: In light of these preliminary data, it is conceivable that in cases of reduced stability, such as those occurring with a type IV bone, all means ameliorating the primary stability and accelerating the osseointegration can be utilized. © Medicina Oral S. L. C.I.F.

The effects of surgical preparation techniques and implant macro-geometry on primary stability: an in vitro study / FALISI, GIOVANNI; Severino, Marco; Rastelli, Claudio; Bernardi, Sara; Caruso, Silvia; GALLI, Massimo; LAMAZZA, LUCA; DI PAOLO, Carlo. - In: MEDICINA ORAL, PATOLOGÍA ORAL Y CIRUGÍA BUCAL. - ISSN 1698-4447. - 22:2(2017), pp. e201-e206. [10.4317/medoral.21286]

The effects of surgical preparation techniques and implant macro-geometry on primary stability: an in vitro study

FALISI, GIOVANNI
Primo
;
GALLI, Massimo;LAMAZZA, LUCA
Penultimo
;
DI PAOLO, Carlo
Ultimo
2017

Abstract

Background: The attainment of a good primary stability is a necessary condition to ensure the success of osseointegration in implantology. In type IV cancellous bone, however, it is possible that a reduced primary stability can lead to an increased rate of failure. The aim of this study was therefore to determine, with the help of the resonance frequency (Osstell mentor), which technique of implant site preparation (piezo surgery, conventional, under-preparation, bone compaction, osteodistraction) and macro-geometry is able to improve implant stability in type IV cancellous bone. Material and Methods: 10 pig ribs were prepared with a surgical pre-drilled guide, calibrated for a correct implant positioning. On each rib, 5 implant sites (one for each technique) were prepared. Successively, 50 conical implants (Tekka Global D) were inserted and measured with the resonance frequency to evaluate the primary stability. Data collected were analyzed by analysis of variance (ANOVA) to test whether the Implant Stability Quotient (ISQ) values of the five techniques were significantly different. Results: The results showed that no significant differences among the ISQ values of the five techniques used were found. Also, no significant differences in the macro-geometry of the two types of compared implants were observed. However, the macro-geometry of Tekka implants, characterized by a double condensing thread, seems to provide greater ISQ values than those of single thread implants when using the same technique. Conclusions: In light of these preliminary data, it is conceivable that in cases of reduced stability, such as those occurring with a type IV bone, all means ameliorating the primary stability and accelerating the osseointegration can be utilized. © Medicina Oral S. L. C.I.F.
2017
implant primary stability; implant site preparation; resonance frequency analysis; surgery; otorhinolaryngology2734 pathology and forensic medicine; dentistry (all)
01 Pubblicazione su rivista::01a Articolo in rivista
The effects of surgical preparation techniques and implant macro-geometry on primary stability: an in vitro study / FALISI, GIOVANNI; Severino, Marco; Rastelli, Claudio; Bernardi, Sara; Caruso, Silvia; GALLI, Massimo; LAMAZZA, LUCA; DI PAOLO, Carlo. - In: MEDICINA ORAL, PATOLOGÍA ORAL Y CIRUGÍA BUCAL. - ISSN 1698-4447. - 22:2(2017), pp. e201-e206. [10.4317/medoral.21286]
File allegati a questo prodotto
File Dimensione Formato  
Falisi_Effects_2017.pdf

accesso aperto

Note: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5359712/
Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 425.92 kB
Formato Adobe PDF
425.92 kB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/957458
Citazioni
  • ???jsp.display-item.citation.pmc??? 22
  • Scopus 56
  • ???jsp.display-item.citation.isi??? 38
social impact