The main purpose of this study is to measure the accuracy and the precision of the intraoral optical scanner CS3500® (Carestream Dental LLC, Atlanta, USA) in complex clinical situations as full arch rehabilitations on impl ants. 50 scans of the acrylic resin model were performed by using CS3500® (Carestream Dental LLC, Atlanta, USA) scanner. Each scan was compared with the virtual model derived from scanning with the laboratory scanner Dscan3® (Enhanced Geometry Soluti on, Bologna, Italy) to measure a possible misalignment. The alignment error was found to be 79,6 ( ± 12,87) m. The measurement was taken at the level of 2 distal scan - abutments. The scanner's precision ranges from 24 to 52 m , depending on the dist ance between scan - abutment. CS3500® (Carestream Dental LLC, Atlanta, USA) intraoral scanner has detected a valid device in the execution of complex rehabilitations on implants. His accuracy and precision values fall within the range established in li terature to define acceptable the prosthetic fitting on full arch implant rehabilitations
Accuracy and precision of an intraoral scanner in complex prosthetic rehabilitations: an in vitro study / DE ANGELIS, Francesca; Pignatiello, Giuseppe; Brauner, Edoardo; Piccoli, Luca; Pompa, Giorgio; DI CARLO, Stefano. - In: JOURNAL OF INTERNATIONAL DENTAL AND MEDICAL RESEARCH. - ISSN 1309-100X. - 10:1(2017), pp. 52-58.
Accuracy and precision of an intraoral scanner in complex prosthetic rehabilitations: an in vitro study
DE ANGELIS, FRANCESCAPrimo
;PIGNATIELLO, GIUSEPPE
Secondo
;BRAUNER, EDOARDO;PICCOLI, Luca;POMPA, GiorgioPenultimo
;DI CARLO, StefanoUltimo
2017
Abstract
The main purpose of this study is to measure the accuracy and the precision of the intraoral optical scanner CS3500® (Carestream Dental LLC, Atlanta, USA) in complex clinical situations as full arch rehabilitations on impl ants. 50 scans of the acrylic resin model were performed by using CS3500® (Carestream Dental LLC, Atlanta, USA) scanner. Each scan was compared with the virtual model derived from scanning with the laboratory scanner Dscan3® (Enhanced Geometry Soluti on, Bologna, Italy) to measure a possible misalignment. The alignment error was found to be 79,6 ( ± 12,87) m. The measurement was taken at the level of 2 distal scan - abutments. The scanner's precision ranges from 24 to 52 m , depending on the dist ance between scan - abutment. CS3500® (Carestream Dental LLC, Atlanta, USA) intraoral scanner has detected a valid device in the execution of complex rehabilitations on implants. His accuracy and precision values fall within the range established in li terature to define acceptable the prosthetic fitting on full arch implant rehabilitationsFile | Dimensione | Formato | |
---|---|---|---|
De-Angelis_Accuracy_2017.pdf
accesso aperto
Note: http://www.jidmr.com/journal/contents-of-2017-vol-10-no-1/
Tipologia:
Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza:
Tutti i diritti riservati (All rights reserved)
Dimensione
544.39 kB
Formato
Adobe PDF
|
544.39 kB | Adobe PDF |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.