The device called Cherenkov detector for proton Flux Measurement is going to be installed inside of theprimary vacuum of the Super Proton Synchrotron to monitor a secondary beam produced by the bentcrystal inserted in the proton halo. Test of this detector with 449 MeV electrons was performed at beamtest facility at Frascati. We measure 0.62 p.e. per incoming electron and a resolution of 15% for 100incoming electrons.

Test of full size Cherenkov detector for proton Flux Measurements / Burmistrov, L.; Breton, D.; Cavoto, Gianluca; Chaumat, V.; Collin, J.; Conforti Di Lorenzo, S.; Garattini, M.; Iacoangeli, F.; Jeglot, J.; Maalmi, J.; Montesano, S.; Puill, V.; Rossi, R.; Scandale, W.; Stocchi, A.; Vagnucci, J. F.. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - 787:(2015), pp. 173-175. [10.1016/j.nima.2014.11.089]

Test of full size Cherenkov detector for proton Flux Measurements

CAVOTO, Gianluca
Membro del Collaboration Group
;
Rossi, R.
Membro del Collaboration Group
;
2015

Abstract

The device called Cherenkov detector for proton Flux Measurement is going to be installed inside of theprimary vacuum of the Super Proton Synchrotron to monitor a secondary beam produced by the bentcrystal inserted in the proton halo. Test of this detector with 449 MeV electrons was performed at beamtest facility at Frascati. We measure 0.62 p.e. per incoming electron and a resolution of 15% for 100incoming electrons.
2015
Cherenkov light; Invacuum detectors; Wave-digitizing electronics; Nuclear and High Energy Physics; Instrumentation
01 Pubblicazione su rivista::01a Articolo in rivista
Test of full size Cherenkov detector for proton Flux Measurements / Burmistrov, L.; Breton, D.; Cavoto, Gianluca; Chaumat, V.; Collin, J.; Conforti Di Lorenzo, S.; Garattini, M.; Iacoangeli, F.; Jeglot, J.; Maalmi, J.; Montesano, S.; Puill, V.; Rossi, R.; Scandale, W.; Stocchi, A.; Vagnucci, J. F.. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - 787:(2015), pp. 173-175. [10.1016/j.nima.2014.11.089]
File allegati a questo prodotto
File Dimensione Formato  
Cavoto_Test of full size Cherenkov_2015.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 580.08 kB
Formato Adobe PDF
580.08 kB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/935327
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 6
social impact