We describe a PET device based on a novel method to extract the coordinates of the interaction point of the 511 keV gamma rays from 100 mm long and thin LYSO (Lutetium Yttrium OxyorthoSilicate) scintillator bars, positioned axially in the tomograph. The coordinate along the hit crystal is measured by using a hodoscope of Wave Length Shifting (WLS) plastic strips mounted perpendicularly to each plane of scintillators. As photodetectors, new Geiger mode Avalanche Photo Detectors (G-APDs) with integrated electronics are being used to detect both the hit crystal in a block (x and y coordinates) and the interaction point in the crystal (z coordinate) through the light escaping from the crystal and transmitted to the WLS strips. In this way, the 7 interaction point can be determined with a spatial resolution of few cubic millimeters clown to a minimum deposited energy of about 50 keV, resulting in a volumetric precision very close to the limits imposed by the physics of the positron annihilation. The method allows to increase the detection efficiency without affecting the spatial resolution by adding scintillator planes in the radial direction. A demonstrator scanner, based on two matrices of 8 x 6 LYSO crystals and 312 WLS strips, slotted in between the crystals, is under construction. Preliminary results from the feasibility studies of the various components will be presented.
Development of a High Precision Axial 3-D PET for Brain Imaging / Bolle, E; Braem, A; Casella, C; Chesi, E; Clinthorne, N; Cochran, E; DE LEO, R; Dissertori, G; Djambazovc, L; Honscheid, K; Huh, S; Johnson, I; Joram, C; Kagan, H; Lacasta, C; Lustermann, W; Meddi, Franco; Nappi, E; NESSI TEDALDI, F; Oliver, Jf; Pauss, F; Rafecas, M; Renker, D; Rudge, A; Schinzel, D; Schneider, T; Seguinot, J; Smith, S; Solevi, P; Stapnes, S; Vilardi, I; Weilhammer, P.. - In: NUCLEAR PHYSICS B-PROCEEDINGS SUPPLEMENTS. - ISSN 0920-5632. - STAMPA. - 197:(2009), pp. 19-23. [10.1016/j.nuclphysbps.2009.10.026]
Development of a High Precision Axial 3-D PET for Brain Imaging
MEDDI, Franco;
2009
Abstract
We describe a PET device based on a novel method to extract the coordinates of the interaction point of the 511 keV gamma rays from 100 mm long and thin LYSO (Lutetium Yttrium OxyorthoSilicate) scintillator bars, positioned axially in the tomograph. The coordinate along the hit crystal is measured by using a hodoscope of Wave Length Shifting (WLS) plastic strips mounted perpendicularly to each plane of scintillators. As photodetectors, new Geiger mode Avalanche Photo Detectors (G-APDs) with integrated electronics are being used to detect both the hit crystal in a block (x and y coordinates) and the interaction point in the crystal (z coordinate) through the light escaping from the crystal and transmitted to the WLS strips. In this way, the 7 interaction point can be determined with a spatial resolution of few cubic millimeters clown to a minimum deposited energy of about 50 keV, resulting in a volumetric precision very close to the limits imposed by the physics of the positron annihilation. The method allows to increase the detection efficiency without affecting the spatial resolution by adding scintillator planes in the radial direction. A demonstrator scanner, based on two matrices of 8 x 6 LYSO crystals and 312 WLS strips, slotted in between the crystals, is under construction. Preliminary results from the feasibility studies of the various components will be presented.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.