Goal: This study aims at a systematic assessment of five computational models of a birdcage coil for magnetic resonance imaging (MRI) with respect to accuracy and computational cost. Methods: The models were implemented using the same geometrical model and numerical algorithm, but different driving methods (i.e., coil “defeaturing”). The defeatured models were labeled as: specific (S2), generic (G32, G16), and hybrid (H16, H16fr-forced). The accuracy of the models was evaluated using the “symmetric mean absolute percentage error” (“SMAPE”), by comparison with measurements in terms of frequency response, as well as electric (||E⃗||) and magnetic (||B⃗||) field magnitude. Results: All the models computed the ||B⃗|| within 35% of the measurements, only the S2, G32, and H16 were able to accurately model the ||E⃗|| inside the phantom with a maximum SMAPE of 16%. Outside the phantom, only the S2 showed a SMAPE lower than 11%. Conclusions: Results showed that assessing the accuracy of ||B⃗|| based only on comparison along the central longitudinal line of the coil can be misleading. Generic or hybrid coils - when properly modeling the currents along the rings/rungs - were sufficient to accurately reproduce the fields inside a phantom while a specific model was needed to accurately model ||E⃗|| in the space between coil and phantom. Significance: Computational modeling of birdcage body coils is extensively used in the evaluation of radiofrequency-induced heating during MRI. Experimental validation of numerical models is needed to determine if a model is an accurate representation of a physical coil.

Assessing the electromagnetic fields generated by a radiofrequency MRI body coil at 64 MHz. Defeaturing versus accuracy / Lucano, Elena; Liberti, Micaela; Mendoza, Gonzalo G.; Lloyd, Tom; Iacono, Maria Ida; Apollonio, Francesca; Wedan, Steve; Kainz, Wolfgang; Angelone, Leonardo M.. - In: IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING. - ISSN 0018-9294. - ELETTRONICO. - 63:8(2016), pp. 1591-1601. [10.1109/TBME.2015.2506680]

Assessing the electromagnetic fields generated by a radiofrequency MRI body coil at 64 MHz. Defeaturing versus accuracy

LUCANO, ELENA;LIBERTI, Micaela;APOLLONIO, Francesca;
2016

Abstract

Goal: This study aims at a systematic assessment of five computational models of a birdcage coil for magnetic resonance imaging (MRI) with respect to accuracy and computational cost. Methods: The models were implemented using the same geometrical model and numerical algorithm, but different driving methods (i.e., coil “defeaturing”). The defeatured models were labeled as: specific (S2), generic (G32, G16), and hybrid (H16, H16fr-forced). The accuracy of the models was evaluated using the “symmetric mean absolute percentage error” (“SMAPE”), by comparison with measurements in terms of frequency response, as well as electric (||E⃗||) and magnetic (||B⃗||) field magnitude. Results: All the models computed the ||B⃗|| within 35% of the measurements, only the S2, G32, and H16 were able to accurately model the ||E⃗|| inside the phantom with a maximum SMAPE of 16%. Outside the phantom, only the S2 showed a SMAPE lower than 11%. Conclusions: Results showed that assessing the accuracy of ||B⃗|| based only on comparison along the central longitudinal line of the coil can be misleading. Generic or hybrid coils - when properly modeling the currents along the rings/rungs - were sufficient to accurately reproduce the fields inside a phantom while a specific model was needed to accurately model ||E⃗|| in the space between coil and phantom. Significance: Computational modeling of birdcage body coils is extensively used in the evaluation of radiofrequency-induced heating during MRI. Experimental validation of numerical models is needed to determine if a model is an accurate representation of a physical coil.
2016
Field probes; finite-difference time domain (FDTD); safety; symmetric mean absolute percentage error (SMAPE); validation; Biomedical Engineering
01 Pubblicazione su rivista::01a Articolo in rivista
Assessing the electromagnetic fields generated by a radiofrequency MRI body coil at 64 MHz. Defeaturing versus accuracy / Lucano, Elena; Liberti, Micaela; Mendoza, Gonzalo G.; Lloyd, Tom; Iacono, Maria Ida; Apollonio, Francesca; Wedan, Steve; Kainz, Wolfgang; Angelone, Leonardo M.. - In: IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING. - ISSN 0018-9294. - ELETTRONICO. - 63:8(2016), pp. 1591-1601. [10.1109/TBME.2015.2506680]
File allegati a questo prodotto
File Dimensione Formato  
Lucano_Assessing_2016.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 1.62 MB
Formato Adobe PDF
1.62 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/911147
Citazioni
  • ???jsp.display-item.citation.pmc??? 12
  • Scopus 32
  • ???jsp.display-item.citation.isi??? 28
social impact