OBJECTIVE: The occurrence of bacterial leakage in the internal surface of implants, through implant-abutment interface (IAI), is one of the parameters for analyzing the fabrication quality of the connections. The aim of this in vitro study is to evaluate two different types of implant-abutment connections: the screwed connection (Group 1) and the cemented connection (Group 2), analyzing the permeability of the IAI to bacterial colonization, using human saliva as culture medium. PATIENTS AND METHODS: A total of twelve implants were tested, six in each experimental group. Five healthy patients were enrolled in this study. Two milliliters of non-stimulated saliva were collected from each subject and mixed in a test tube. After 14 days of incubation of the bacteria sample in the implant fixtures, a PCR-Real Time analysis was performed. Fisher's exact test was used to compare the proportions of implant-abutment assembled structures detected with bacterial leakage. Differences in the bacterial counts of the two groups were compared using the Mann-Whitney U test. A p value < 0.05 was considered significant. RESULTS: The results showed a decreased stability with the screwed implant-abutment connections compared to the cemented implant-abutment connections. A mean total bacterial count of 1.2E+07 (± 0.25E+07) for Group 1 and of 7.2E+04 (± 14.4E+04) for Group 2 was found, with a high level of significance, p = .0001. CONCLUSIONS: Within the limitations of this study it can be concluded that bacterial species from human saliva may penetrate along the implant-abutment interface in both connections, however the cemented connection implants showed the lowest amount of bacterial colonization.

Salivary bacterial leakage into implant-abutment connections: preliminary results of an in vitro study / Mencio, F; Papi, P; Di Carlo, S; Pompa, G. - In: EUROPEAN REVIEW FOR MEDICAL AND PHARMACOLOGICAL SCIENCES. - ISSN 2284-0729. - STAMPA. - 20:12(2016), pp. 2476-2483.

Salivary bacterial leakage into implant-abutment connections: preliminary results of an in vitro study

Mencio, F
Primo
;
Papi, P
Secondo
;
Di Carlo, S
Penultimo
;
Pompa, G
Ultimo
2016

Abstract

OBJECTIVE: The occurrence of bacterial leakage in the internal surface of implants, through implant-abutment interface (IAI), is one of the parameters for analyzing the fabrication quality of the connections. The aim of this in vitro study is to evaluate two different types of implant-abutment connections: the screwed connection (Group 1) and the cemented connection (Group 2), analyzing the permeability of the IAI to bacterial colonization, using human saliva as culture medium. PATIENTS AND METHODS: A total of twelve implants were tested, six in each experimental group. Five healthy patients were enrolled in this study. Two milliliters of non-stimulated saliva were collected from each subject and mixed in a test tube. After 14 days of incubation of the bacteria sample in the implant fixtures, a PCR-Real Time analysis was performed. Fisher's exact test was used to compare the proportions of implant-abutment assembled structures detected with bacterial leakage. Differences in the bacterial counts of the two groups were compared using the Mann-Whitney U test. A p value < 0.05 was considered significant. RESULTS: The results showed a decreased stability with the screwed implant-abutment connections compared to the cemented implant-abutment connections. A mean total bacterial count of 1.2E+07 (± 0.25E+07) for Group 1 and of 7.2E+04 (± 14.4E+04) for Group 2 was found, with a high level of significance, p = .0001. CONCLUSIONS: Within the limitations of this study it can be concluded that bacterial species from human saliva may penetrate along the implant-abutment interface in both connections, however the cemented connection implants showed the lowest amount of bacterial colonization.
2016
implant-abutment interface; screwed connection; cemented connection; peri-implant; bacterial adhesion
01 Pubblicazione su rivista::01a Articolo in rivista
Salivary bacterial leakage into implant-abutment connections: preliminary results of an in vitro study / Mencio, F; Papi, P; Di Carlo, S; Pompa, G. - In: EUROPEAN REVIEW FOR MEDICAL AND PHARMACOLOGICAL SCIENCES. - ISSN 2284-0729. - STAMPA. - 20:12(2016), pp. 2476-2483.
File allegati a questo prodotto
File Dimensione Formato  
Mencio_Salivary_2016.pdf

accesso aperto

Note: https://www.europeanreview.org/article/11000
Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 2.13 MB
Formato Adobe PDF
2.13 MB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/910979
Citazioni
  • ???jsp.display-item.citation.pmc??? 8
  • Scopus 17
  • ???jsp.display-item.citation.isi??? 17
social impact