High-flow nasal cannula (HFNC) is frequently used to treat respiratory distress in infants and children because of its beneficial effects on alveolar ventilation and respiratory mechanics. Setting an adequate flow rate that meets a patient's peak inspiratory flow (PIF) is thus crucially important to achieve such effects. HFNC flow rate is typically set at 1 L/min/kg +1 as suggested by the manufacturer and increased to 2 L/min/kg according to the degree of respiratory distress. However, whether this empirical flow setting actually meets a patient's PIF has not yet been investigated. In this study, we implemented our previously described respiratory mechanics monitoring system (MAES) with a new spirometry function (NSS) that allows for a simultaneous visualization of the flow tracings of HFNC and the patient's spontaneous breathing. We tested the ability of NSS-MAES to determine the adequacy of empirically set flow rates of 1 L/min/kg +1 or 2 L/min/kg on 9 infants with respiratory distress receiving HFNC. HFNC flow rate was considered adequate if its tracing was just above the patient's respiratory flow. In patients in whom 1 L/kg/min +1 was inadequate, we used NSS-MAES to identify the adequate flow by raising the HFNC flow until it reached the patient's PIF (HFNC-NSS-MAES). We also investigated which flow rate was associated with the maximal decrease of respiratory effort, namely, Pressure Time Product (PTP) and Work of Breathing (WOB). We found that 1 L/min/kg +1, but not 2 L/min/kg was often unable to meet the patient's PIF. In these cases HFNC-NSS-MAES values were around 1.6 L/min/kg. Conversely, HFNC at 2 L/min/kg always exceeded the patient's PIF. All breathing effort indexes tested improved after HFNC treatment with the maximal unloading seen at 2 L/min/kg for PTP and at HFNC-NSS-MAES. © 2016 IEEE.

Optimization of flow setting during high-flow nasal cannula (HFNC) with a new spirometry system / Montecchia, F.; Midulla, Fabio; Moretti, Corrado; Papoff, Paola. - ELETTRONICO. - (2016), pp. 92-97. (Intervento presentato al convegno MEMEA 2016 tenutosi a BENEVENTO) [10.1109/MeMeA.2016.7533714].

Optimization of flow setting during high-flow nasal cannula (HFNC) with a new spirometry system

MIDULLA, Fabio;MORETTI, Corrado;PAPOFF, PAOLA
2016

Abstract

High-flow nasal cannula (HFNC) is frequently used to treat respiratory distress in infants and children because of its beneficial effects on alveolar ventilation and respiratory mechanics. Setting an adequate flow rate that meets a patient's peak inspiratory flow (PIF) is thus crucially important to achieve such effects. HFNC flow rate is typically set at 1 L/min/kg +1 as suggested by the manufacturer and increased to 2 L/min/kg according to the degree of respiratory distress. However, whether this empirical flow setting actually meets a patient's PIF has not yet been investigated. In this study, we implemented our previously described respiratory mechanics monitoring system (MAES) with a new spirometry function (NSS) that allows for a simultaneous visualization of the flow tracings of HFNC and the patient's spontaneous breathing. We tested the ability of NSS-MAES to determine the adequacy of empirically set flow rates of 1 L/min/kg +1 or 2 L/min/kg on 9 infants with respiratory distress receiving HFNC. HFNC flow rate was considered adequate if its tracing was just above the patient's respiratory flow. In patients in whom 1 L/kg/min +1 was inadequate, we used NSS-MAES to identify the adequate flow by raising the HFNC flow until it reached the patient's PIF (HFNC-NSS-MAES). We also investigated which flow rate was associated with the maximal decrease of respiratory effort, namely, Pressure Time Product (PTP) and Work of Breathing (WOB). We found that 1 L/min/kg +1, but not 2 L/min/kg was often unable to meet the patient's PIF. In these cases HFNC-NSS-MAES values were around 1.6 L/min/kg. Conversely, HFNC at 2 L/min/kg always exceeded the patient's PIF. All breathing effort indexes tested improved after HFNC treatment with the maximal unloading seen at 2 L/min/kg for PTP and at HFNC-NSS-MAES. © 2016 IEEE.
2016
MEMEA 2016
biomedical signal processing; biomedical transducers; clinical trials; pediatrics; respiratory mechanics; Signal Processing; Biomedical Engineering; Instrumentation
04 Pubblicazione in atti di convegno::04b Atto di convegno in volume
Optimization of flow setting during high-flow nasal cannula (HFNC) with a new spirometry system / Montecchia, F.; Midulla, Fabio; Moretti, Corrado; Papoff, Paola. - ELETTRONICO. - (2016), pp. 92-97. (Intervento presentato al convegno MEMEA 2016 tenutosi a BENEVENTO) [10.1109/MeMeA.2016.7533714].
File allegati a questo prodotto
File Dimensione Formato  
Montecchia_HFNC_2016.pdf

solo utenti autorizzati

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 2.29 MB
Formato Adobe PDF
2.29 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/906447
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 0
social impact