AIMS:Thrombus formation leading to ischaemic stroke is a major concern after transcathether occlusion of interatrial communications. The aim of our study was to verify if postprocedural transoesophageal echocardiography (TEE) might improve timely detection and management of thrombus formation. METHODS AND RESULTS:We studied 65 patients with patent foramen ovale or atrial septal defect who received an atrial septal closure device. Transoesophageal echocardiography was used to guide the procedure, and was repeated <1 h after the end of the procedure, at 30 day and 6 month follow-up visits. Transoesophageal echocardiography <1 h after procedure revealed a left-sided thrombus attached to the device in three patients. All cases were successfully treated by thrombolytic therapy with full-dose tenecteplase (0.53 mg/kg in 5-10 s) with disappearance of thrombus in 15-60 min. No cerebrovascular event was recorded during a follow-up of 311 +/- 235 days. CONCLUSION:Thrombus formation on interatrial closure devices can occur soon after device implantation and can be timely detected by postprocedural TEE.

Transcathether occlusion of interatrial communications. postprocedural transoesophageal echocardiography allows timely detection and treatment of intracardiac thrombus formation / Pelliccia, Francesco; Cianfrocca, Cinzia; Pasceri, Vincenzo; Granatelli, Antonino; Speciale, Giulio; Roncella, Adriana; Turturo, Maurizio; Richichi, Giuseppe; Pristipino, Christian. - In: EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY. - ISSN 1525-2167. - STAMPA. - 10:3(2009), pp. 439-441. [10.1093/ejechocard/jen306]

Transcathether occlusion of interatrial communications. postprocedural transoesophageal echocardiography allows timely detection and treatment of intracardiac thrombus formation

PELLICCIA, FRANCESCO
;
2009

Abstract

AIMS:Thrombus formation leading to ischaemic stroke is a major concern after transcathether occlusion of interatrial communications. The aim of our study was to verify if postprocedural transoesophageal echocardiography (TEE) might improve timely detection and management of thrombus formation. METHODS AND RESULTS:We studied 65 patients with patent foramen ovale or atrial septal defect who received an atrial septal closure device. Transoesophageal echocardiography was used to guide the procedure, and was repeated <1 h after the end of the procedure, at 30 day and 6 month follow-up visits. Transoesophageal echocardiography <1 h after procedure revealed a left-sided thrombus attached to the device in three patients. All cases were successfully treated by thrombolytic therapy with full-dose tenecteplase (0.53 mg/kg in 5-10 s) with disappearance of thrombus in 15-60 min. No cerebrovascular event was recorded during a follow-up of 311 +/- 235 days. CONCLUSION:Thrombus formation on interatrial closure devices can occur soon after device implantation and can be timely detected by postprocedural TEE.
2009
atrial septal communication; thrombolysis; thrombus; transcatheter occlusion; transoesophageal echocardiography; adult; aged; aged, 80 and over; echocardiography, transesophageal; female; fibrinolytic agents; follow-up studies; foramen ovale, patent; heart diseases; humans; male; middle aged; stroke; thrombosis; tissue plasminogen activator; ultrasonography, interventional; young adult; cardiology and cardiovascular medicine; radiology, nuclear medicine and imaging
01 Pubblicazione su rivista::01a Articolo in rivista
Transcathether occlusion of interatrial communications. postprocedural transoesophageal echocardiography allows timely detection and treatment of intracardiac thrombus formation / Pelliccia, Francesco; Cianfrocca, Cinzia; Pasceri, Vincenzo; Granatelli, Antonino; Speciale, Giulio; Roncella, Adriana; Turturo, Maurizio; Richichi, Giuseppe; Pristipino, Christian. - In: EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY. - ISSN 1525-2167. - STAMPA. - 10:3(2009), pp. 439-441. [10.1093/ejechocard/jen306]
File allegati a questo prodotto
File Dimensione Formato  
Pelliccia_Transcathether_2009.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 135.93 kB
Formato Adobe PDF
135.93 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/894184
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 3
social impact