OBJECTIVES: To compare single-energy (SECT) and dual-energy (DECT) abdominal CT examinations in matched patient cohorts regarding differences in radiation dose and image quality performed with second- and third-generation dual-source CT (DSCT). METHODS: We retrospectively analysed 200 patients (100 male, 100 female; mean age 61.2 ± 13.5 years, mean body mass index 27.5 ± 3.8 kg/m2) equally divided into four groups matched by gender and body mass index, who had undergone portal venous phase abdominal CT with second-generation (group A, 120-kV-SECT; group B, 80/140-kV-DECT) and third-generation DSCT (group C, 100-kV-SECT; group D, 90/150-kV-DECT). The radiation dose was normalised for 40-cm scan length. Dose-independent figure-of-merit (FOM) contrast-to-noise ratios (CNRs) were calculated for various organs and vessels. Subjective overall image quality and reader confidence were assessed. RESULTS: The effective normalised radiation dose was significantly lower (P < 0.001) in groups C (6.2 ± 2.0 mSv) and D (5.3 ± 1.9 mSv, P = 0.103) compared to groups A (8.8 ± 2.3 mSv) and B (9.7 ± 2.4 mSv, P = 0.102). Dose-independent FOM-CNR peaked for liver, kidney, and portal vein measurements (all P ≤ 0.0285) in group D. Subjective image quality and reader confidence were consistently rated as excellent in all groups (all ≥1.53 out of 5). CONCLUSIONS: With both DSCT generations, abdominal DECT can be routinely performed without radiation dose penalty compared to SECT, while third-generation DSCT shows improved dose efficiency.

Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT / Wichmann, Julian L; Hardie, Andrew D; Schoepf, U. Joseph; Felmly, Lloyd M; Perry, Jonathan D; Varga Szemes, Akos; Mangold, Stefanie; Caruso, Damiano; Canstein, Christian; Vogl, Thomas J; DE CECCO, CARLO NICOLA. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - ELETTRONICO. - 27:(2017), pp. 642-650. [10.1007/s00330-016-4383-6]

Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT

CARUSO, DAMIANO;DE CECCO, CARLO NICOLA
2017

Abstract

OBJECTIVES: To compare single-energy (SECT) and dual-energy (DECT) abdominal CT examinations in matched patient cohorts regarding differences in radiation dose and image quality performed with second- and third-generation dual-source CT (DSCT). METHODS: We retrospectively analysed 200 patients (100 male, 100 female; mean age 61.2 ± 13.5 years, mean body mass index 27.5 ± 3.8 kg/m2) equally divided into four groups matched by gender and body mass index, who had undergone portal venous phase abdominal CT with second-generation (group A, 120-kV-SECT; group B, 80/140-kV-DECT) and third-generation DSCT (group C, 100-kV-SECT; group D, 90/150-kV-DECT). The radiation dose was normalised for 40-cm scan length. Dose-independent figure-of-merit (FOM) contrast-to-noise ratios (CNRs) were calculated for various organs and vessels. Subjective overall image quality and reader confidence were assessed. RESULTS: The effective normalised radiation dose was significantly lower (P < 0.001) in groups C (6.2 ± 2.0 mSv) and D (5.3 ± 1.9 mSv, P = 0.103) compared to groups A (8.8 ± 2.3 mSv) and B (9.7 ± 2.4 mSv, P = 0.102). Dose-independent FOM-CNR peaked for liver, kidney, and portal vein measurements (all P ≤ 0.0285) in group D. Subjective image quality and reader confidence were consistently rated as excellent in all groups (all ≥1.53 out of 5). CONCLUSIONS: With both DSCT generations, abdominal DECT can be routinely performed without radiation dose penalty compared to SECT, while third-generation DSCT shows improved dose efficiency.
2017
abdomen; diagnostic imaging; multidetector computed tomography; radiation dosage; retrospective study
01 Pubblicazione su rivista::01a Articolo in rivista
Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT / Wichmann, Julian L; Hardie, Andrew D; Schoepf, U. Joseph; Felmly, Lloyd M; Perry, Jonathan D; Varga Szemes, Akos; Mangold, Stefanie; Caruso, Damiano; Canstein, Christian; Vogl, Thomas J; DE CECCO, CARLO NICOLA. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - ELETTRONICO. - 27:(2017), pp. 642-650. [10.1007/s00330-016-4383-6]
File allegati a questo prodotto
File Dimensione Formato  
Wichmann_Abdomen_2017.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 1.22 MB
Formato Adobe PDF
1.22 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/868302
Citazioni
  • ???jsp.display-item.citation.pmc??? 23
  • Scopus 87
  • ???jsp.display-item.citation.isi??? 78
social impact