The Optoelectronic Systems (OS) are largely used in gait analysis to evaluate the motor performances of healthy subjects and patients. The accuracy of marker trajectories reconstruction depends on several aspects: the number of cameras, the dimension and position of the calibration volume, and the chosen calibration procedure. In this paper we propose a methodology to evaluate the eects of the mentioned sources of error on the reconstruction of marker trajectories. The novel contribution of the present work consists in the dimension of the tested calibration volumes, which is comparable with the ones normally used in gait analysis; in addition, to simulate trajectories during clinical gait analysis, we provide non-default paths for markers as inputs. Several calibration procedures are implemented and the same trial is processed with each calibration le, also considering dierent cameras congurations. The RMSEs between the measured trajectories and the optimal ones are calculated for each comparison. To investigate the signicant dierences between the computed indices, an ANOVA analysis is implemented. The RMSE is sensible to the variations of the considered calibration volume and the camera congurations and it is always inferior to 43 mm.

Technical quality assessment of an optoelectronic system for movement analysis / DI MARCO, 102. R.; Rossi, S.; Patané, F.; Cappa, P.. - In: JOURNAL OF PHYSICS. CONFERENCE SERIES. - ISSN 1742-6588. - STAMPA. - 588:(2015). [10.1088/1742-6596/588/1/012030]

Technical quality assessment of an optoelectronic system for movement analysis

S. ROSSI;F. PATANÉ;P. CAPPA
2015

Abstract

The Optoelectronic Systems (OS) are largely used in gait analysis to evaluate the motor performances of healthy subjects and patients. The accuracy of marker trajectories reconstruction depends on several aspects: the number of cameras, the dimension and position of the calibration volume, and the chosen calibration procedure. In this paper we propose a methodology to evaluate the eects of the mentioned sources of error on the reconstruction of marker trajectories. The novel contribution of the present work consists in the dimension of the tested calibration volumes, which is comparable with the ones normally used in gait analysis; in addition, to simulate trajectories during clinical gait analysis, we provide non-default paths for markers as inputs. Several calibration procedures are implemented and the same trial is processed with each calibration le, also considering dierent cameras congurations. The RMSEs between the measured trajectories and the optimal ones are calculated for each comparison. To investigate the signicant dierences between the computed indices, an ANOVA analysis is implemented. The RMSE is sensible to the variations of the considered calibration volume and the camera congurations and it is always inferior to 43 mm.
2015
calibration; cameras; gait analysis; optoelectronic devices; trajectories
01 Pubblicazione su rivista::01a Articolo in rivista
Technical quality assessment of an optoelectronic system for movement analysis / DI MARCO, 102. R.; Rossi, S.; Patané, F.; Cappa, P.. - In: JOURNAL OF PHYSICS. CONFERENCE SERIES. - ISSN 1742-6588. - STAMPA. - 588:(2015). [10.1088/1742-6596/588/1/012030]
File allegati a questo prodotto
File Dimensione Formato  
Di Marco_Technical-quality_2015.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 1.66 MB
Formato Adobe PDF
1.66 MB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/784363
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 4
social impact