Background. Overheating during implant site preparation negatively affects the osseointegration process as well the final outcome of implant rehabilitations. Piezoelectric techniques seem to provide to a gentle implant preparation although few scientific reports have investigated the heat generation and its underlying factors. Purpose. To investigate, through a proper methodological approach, the main factors influencing temperature rise during piezoelectric implant site preparation. Materials and methods. Different piezoelectric tips (IM1s, IM2, P2-3, IM3, Mectron Medical Technology, Carasco, Italy) have been tested. The experimental set-up consisted in a mechanical positioning device equipped with a load cell and a fluoroptic thermometer. Results. The first tip of the sequence (IM1s) generated the highest temperature increasing (ΔT). The diamond tips (IM1s and P2-3) determined higher ΔT values than the smooth tips (IM2 and IM3). Further tests with IM1s suggested that the temperature

Identification of possible factors influencing temperatures elevation nduring implant site preparation with piezoelectic technique / Lamazza, Luca; Laurito, Domenica; Lollobrigida, M.; Brugnoletti, Orlando; Garreffa, G.; DE BIASE, Alberto. - In: ANNALI DI STOMATOLOGIA. - ISSN 1971-1441. - V:4(2014), pp. 115-122.

Identification of possible factors influencing temperatures elevation nduring implant site preparation with piezoelectic technique

LAMAZZA, LUCA;LAURITO, DOMENICA;M. Lollobrigida;BRUGNOLETTI, Orlando;DE BIASE, Alberto
2014

Abstract

Background. Overheating during implant site preparation negatively affects the osseointegration process as well the final outcome of implant rehabilitations. Piezoelectric techniques seem to provide to a gentle implant preparation although few scientific reports have investigated the heat generation and its underlying factors. Purpose. To investigate, through a proper methodological approach, the main factors influencing temperature rise during piezoelectric implant site preparation. Materials and methods. Different piezoelectric tips (IM1s, IM2, P2-3, IM3, Mectron Medical Technology, Carasco, Italy) have been tested. The experimental set-up consisted in a mechanical positioning device equipped with a load cell and a fluoroptic thermometer. Results. The first tip of the sequence (IM1s) generated the highest temperature increasing (ΔT). The diamond tips (IM1s and P2-3) determined higher ΔT values than the smooth tips (IM2 and IM3). Further tests with IM1s suggested that the temperature
2014
01 Pubblicazione su rivista::01a Articolo in rivista
Identification of possible factors influencing temperatures elevation nduring implant site preparation with piezoelectic technique / Lamazza, Luca; Laurito, Domenica; Lollobrigida, M.; Brugnoletti, Orlando; Garreffa, G.; DE BIASE, Alberto. - In: ANNALI DI STOMATOLOGIA. - ISSN 1971-1441. - V:4(2014), pp. 115-122.
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/763896
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact