To evaluate the effect of automated tube voltage selection on radiation dose and image quality at cardiovascular CT angiography (CTA).We retrospectively analysed paired studies in 72 patients (41 male, 60.5 ± 16.5 years), who had undergone CTA acquisitions of the heart or aorta both before and after the implementation of an automated x-ray tube voltage selection algorithm (ATVS). All other parameters were kept identical between the two acquisitions. Subjective image quality (IQ) was rated and objective IQ was measured by image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and figure of merit (FOM). Image quality parameters and effective dose were compared between acquisitions.Overall subjective image quality improved with the percentage of cases scored as adequate or higher increasing from 79 \% to 92 \% after implementation of ATVS (P = 0.03). SNR (14.1 ± 5.9, 15.7 ± 6.1, P = 0.009), CNR (11.6 ± 5.3, 13.2 ± 5.6, P = 0.011), and FOM (19.9 ± 23.3, 43.8 ± 51.1, P < 0.001) were significantly higher after implementation of ATVS. Mean image noise (24.1 ± 8.4 HU, 22.7 ± 7.1 HU, P = 0.048) and mean effective dose (10.6 ± 5.9 mSv, 8.8 ± 5.0 mSv, P = 0.003) were significantly lower after implementation of ATVS.Automated tube voltage selection can operator-independently optimize cardiovascular CTA image acquisition parameters with improved image quality at reduced dose.• Automatic tube voltage selection optimizes tube voltage for each individual patient. • In this population, overall radiation dose decreased while image quality improved. • This tool may become valuable for improving dose/quality ratio.

Reduced radiation dose and improved image quality at cardiovascular CT angiography by automated attenuation-based tube voltage selection: intra-individual comparison / A. W., Krazinski; F. G., Meinel; U. J., Schoepf; J. R., Silverman; C., Canstein; DE CECCO, CARLO NICOLA; L. L., Geyer. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - 24:(2014), pp. 2677-2684. [10.1007/s00330-014-3312-9]

Reduced radiation dose and improved image quality at cardiovascular CT angiography by automated attenuation-based tube voltage selection: intra-individual comparison.

DE CECCO, CARLO NICOLA;
2014

Abstract

To evaluate the effect of automated tube voltage selection on radiation dose and image quality at cardiovascular CT angiography (CTA).We retrospectively analysed paired studies in 72 patients (41 male, 60.5 ± 16.5 years), who had undergone CTA acquisitions of the heart or aorta both before and after the implementation of an automated x-ray tube voltage selection algorithm (ATVS). All other parameters were kept identical between the two acquisitions. Subjective image quality (IQ) was rated and objective IQ was measured by image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and figure of merit (FOM). Image quality parameters and effective dose were compared between acquisitions.Overall subjective image quality improved with the percentage of cases scored as adequate or higher increasing from 79 \% to 92 \% after implementation of ATVS (P = 0.03). SNR (14.1 ± 5.9, 15.7 ± 6.1, P = 0.009), CNR (11.6 ± 5.3, 13.2 ± 5.6, P = 0.011), and FOM (19.9 ± 23.3, 43.8 ± 51.1, P < 0.001) were significantly higher after implementation of ATVS. Mean image noise (24.1 ± 8.4 HU, 22.7 ± 7.1 HU, P = 0.048) and mean effective dose (10.6 ± 5.9 mSv, 8.8 ± 5.0 mSv, P = 0.003) were significantly lower after implementation of ATVS.Automated tube voltage selection can operator-independently optimize cardiovascular CTA image acquisition parameters with improved image quality at reduced dose.• Automatic tube voltage selection optimizes tube voltage for each individual patient. • In this population, overall radiation dose decreased while image quality improved. • This tool may become valuable for improving dose/quality ratio.
2014
01 Pubblicazione su rivista::01a Articolo in rivista
Reduced radiation dose and improved image quality at cardiovascular CT angiography by automated attenuation-based tube voltage selection: intra-individual comparison / A. W., Krazinski; F. G., Meinel; U. J., Schoepf; J. R., Silverman; C., Canstein; DE CECCO, CARLO NICOLA; L. L., Geyer. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - 24:(2014), pp. 2677-2684. [10.1007/s00330-014-3312-9]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/758931
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 7
  • Scopus 31
  • ???jsp.display-item.citation.isi??? 28
social impact