Using Goldmann applanation tonometry (GAT) as a gold standard, we evaluated the accuracy of Nidek NT-4000 pneumotonometry (NPT) in adults without corneal disease. Bland and Altman analysis of serial intra-ocular pressures (IOPs) measured with NPT and GAT in 10 healthy subjects revealed that the repeatability coefficients for the two methods were similar. NPT, GAT and ultrasonic pachymetry were then performed in 100 patients. Bland and Altman analysis showed that NPT yielded significantly higher readings than GAT [mean biases for right and left eye measurements were 1.37 mmHg (95% limits of agreement: -3.02-5.76) and 1.17 mmHg (95% limits of agreement: -2.76-5.11) respectively] and was more affected by corneal thickness variations. For detection of IOPs >= 21 mmHg, NIPT displayed very high sensitivity (0.90) and good specificity (0.95). NPT may be useful in screening and clinical settings but borderline-high IOP readings should be confirmed with GAT.

Validity and limitations of the Nidek NT-4000 non-contact tonometer: a clinical study / Federico, Regine; Scuderi, Gianluca; Massimo, Cesareo; Federico, Ricci; Claudio, Cedrone; Carlo, Nucci. - In: OPHTHALMIC AND PHYSIOLOGICAL OPTICS. - ISSN 0275-5408. - 26:1(2006), pp. 33-39. [10.1111/j.1475-1313.2005.00341.x]

Validity and limitations of the Nidek NT-4000 non-contact tonometer: a clinical study

SCUDERI, GIANLUCA;
2006

Abstract

Using Goldmann applanation tonometry (GAT) as a gold standard, we evaluated the accuracy of Nidek NT-4000 pneumotonometry (NPT) in adults without corneal disease. Bland and Altman analysis of serial intra-ocular pressures (IOPs) measured with NPT and GAT in 10 healthy subjects revealed that the repeatability coefficients for the two methods were similar. NPT, GAT and ultrasonic pachymetry were then performed in 100 patients. Bland and Altman analysis showed that NPT yielded significantly higher readings than GAT [mean biases for right and left eye measurements were 1.37 mmHg (95% limits of agreement: -3.02-5.76) and 1.17 mmHg (95% limits of agreement: -2.76-5.11) respectively] and was more affected by corneal thickness variations. For detection of IOPs >= 21 mmHg, NIPT displayed very high sensitivity (0.90) and good specificity (0.95). NPT may be useful in screening and clinical settings but borderline-high IOP readings should be confirmed with GAT.
2006
central corneal thickness; limits of agreement; nidek nt-4000; pneumotonometry; repeatability coefficient; sensitivity; specificity
01 Pubblicazione su rivista::01a Articolo in rivista
Validity and limitations of the Nidek NT-4000 non-contact tonometer: a clinical study / Federico, Regine; Scuderi, Gianluca; Massimo, Cesareo; Federico, Ricci; Claudio, Cedrone; Carlo, Nucci. - In: OPHTHALMIC AND PHYSIOLOGICAL OPTICS. - ISSN 0275-5408. - 26:1(2006), pp. 33-39. [10.1111/j.1475-1313.2005.00341.x]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/75693
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 6
  • Scopus 27
  • ???jsp.display-item.citation.isi??? 20
social impact