Objective: To estimate the reliability of the Bone Conduction-HeadBand (BC-HB) test for predicting the postoperative functional outcome of a round-window (RW) vibroplasty. Study Design: Within-subject comparison of the functional results of the BC-HB test, which is routinely used for the preoperative evaluation of a bone-conduction transducer, with an active middle ear implant (AMEI) placed onto the round window. Setting: Tertiary referral university hospital center. Patients: Seven patients with similar anatomic (absent stapes superstructure) and functional (moderate, mixed hearing loss) sequelae from open tympanoplasty technique. Intervention: All subjects underwent preoperative audiologic assessment with the BC-HB. Subsequently, all subjects underwent surgical placement of an AMEI onto the round window. Main Outcome Measure: Pure tone and speech audiometry in quiet and noise were assessed. Additionally, evaluation of specific satisfactory targets was performed using the Client Oriented Scale of Improvement. Results: Pure tone and speech audiometry in quiet established that both devices had very similar performance and provided remarkable improvement compared with the unaided condition. However, high-frequency gain and speech audiometry in noise demonstrated better performance with RW-AMEI. Conclusion: In patients presenting with mixed hearing loss as a sequela from middle ear surgery, the preoperative BC-HB test may be helpful in predicting the final functional outcome and patient satisfaction with RW-AMEI.

Is the Bone-Conduction Headband Test Useful for Predicting the Functional Outcome of a Round Window Active Middle Ear Implant? / Monini, Simonetta; Filippi, Chiara; Atturo, Francesca; Barbara, Maurizio. - In: OTOLOGY & NEUROTOLOGY. - ISSN 1531-7129. - STAMPA. - 34:7(2013), pp. 1329-1335. [10.1097/mao.0b013e3182953100]

Is the Bone-Conduction Headband Test Useful for Predicting the Functional Outcome of a Round Window Active Middle Ear Implant?

MONINI, Simonetta;FILIPPI, CHIARA;ATTURO, FRANCESCA;BARBARA, Maurizio
2013

Abstract

Objective: To estimate the reliability of the Bone Conduction-HeadBand (BC-HB) test for predicting the postoperative functional outcome of a round-window (RW) vibroplasty. Study Design: Within-subject comparison of the functional results of the BC-HB test, which is routinely used for the preoperative evaluation of a bone-conduction transducer, with an active middle ear implant (AMEI) placed onto the round window. Setting: Tertiary referral university hospital center. Patients: Seven patients with similar anatomic (absent stapes superstructure) and functional (moderate, mixed hearing loss) sequelae from open tympanoplasty technique. Intervention: All subjects underwent preoperative audiologic assessment with the BC-HB. Subsequently, all subjects underwent surgical placement of an AMEI onto the round window. Main Outcome Measure: Pure tone and speech audiometry in quiet and noise were assessed. Additionally, evaluation of specific satisfactory targets was performed using the Client Oriented Scale of Improvement. Results: Pure tone and speech audiometry in quiet established that both devices had very similar performance and provided remarkable improvement compared with the unaided condition. However, high-frequency gain and speech audiometry in noise demonstrated better performance with RW-AMEI. Conclusion: In patients presenting with mixed hearing loss as a sequela from middle ear surgery, the preoperative BC-HB test may be helpful in predicting the final functional outcome and patient satisfaction with RW-AMEI.
2013
pure tone audiometry.; headband; speech audiometry; open tympanoplasty; bone anchored hearing device; client-oriented scale of improvement; bone anchored implant; mixed hearing loss; vibroplasty; active middle ear implant
01 Pubblicazione su rivista::01a Articolo in rivista
Is the Bone-Conduction Headband Test Useful for Predicting the Functional Outcome of a Round Window Active Middle Ear Implant? / Monini, Simonetta; Filippi, Chiara; Atturo, Francesca; Barbara, Maurizio. - In: OTOLOGY & NEUROTOLOGY. - ISSN 1531-7129. - STAMPA. - 34:7(2013), pp. 1329-1335. [10.1097/mao.0b013e3182953100]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/609201
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 8
social impact