Fatigue behavior of hip prostheses is usually assessed experimentally, through the implementation of mechanical tests according to International Technical Standards. Based on FE elastic-linear simulations, this study aims at investigating the stress resulting from loading conditions which are slightly different from the reference condition indicated in the ISO7206-4:2010 but which are reasonable from both a clinical and biomechanical point of view. Main results are reported and suggestions delivered for the assessment of risky conditions.
Fatigue behaviour of artificial femoral stems: the role of Regulations and FE analysis / Andreaus, Ugo; C., Giacomozzi; M., Rogante; I., Campioni. - ELETTRONICO. - 1:(2012), pp. 1-3. (Intervento presentato al convegno The National Congress of Italian Group of Bioengineering tenutosi a Roma nel June 26-29).
Fatigue behaviour of artificial femoral stems: the role of Regulations and FE analysis
ANDREAUS, Ugo;
2012
Abstract
Fatigue behavior of hip prostheses is usually assessed experimentally, through the implementation of mechanical tests according to International Technical Standards. Based on FE elastic-linear simulations, this study aims at investigating the stress resulting from loading conditions which are slightly different from the reference condition indicated in the ISO7206-4:2010 but which are reasonable from both a clinical and biomechanical point of view. Main results are reported and suggestions delivered for the assessment of risky conditions.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.