We report on the treatment of a patient, with severe peri-prosthetic regurgitation, following CoreValve implantation too low into the left ventricle outflow tract, adjusted by 'snaring' prosthesis. This bail-out manoeuvre allows the appropriate positioning of the prosthesis but was complicated by the entrapment of the goose-neck loop in the valve delivery anchor and was solved successfully with a second different device.
CoreValve repositioning complicated by entrapment of snare in the valve delivery anchor / Rocco Edoardo, Stio; Sardella, Gennaro; Mauro, Pennacchi; Gian Paolo, Ussia; Fedele, Francesco. - In: JOURNAL OF CARDIOVASCULAR MEDICINE. - ISSN 1558-2027. - (2013), p. 1. [10.2459/jcm.0b013e32835dbc92]
CoreValve repositioning complicated by entrapment of snare in the valve delivery anchor
SARDELLA, Gennaro;FEDELE, Francesco
2013
Abstract
We report on the treatment of a patient, with severe peri-prosthetic regurgitation, following CoreValve implantation too low into the left ventricle outflow tract, adjusted by 'snaring' prosthesis. This bail-out manoeuvre allows the appropriate positioning of the prosthesis but was complicated by the entrapment of the goose-neck loop in the valve delivery anchor and was solved successfully with a second different device.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.