The purpose of this study was to evaluate the diagnostic performance of vertebral fracture assessment (VFA) using the Lunar iDXA scanner. Conventional spinal radiographs and images acquired by dual-energy X-ray absorptiometry (DXA) of 350 subjects (269 females, 81 males) were evaluated by two different readers. We visualized 4,476/4,550 (98.4 %) vertebrae from T4 to L4 on VFA images compared to 4,535/4,550 (99.7 %) on radiographs. Among the visualized vertebrae, 205/4,535 (4.5 %) and 190/4,476 (4.2 %) were identified as nonfracture deformities by reading of radiographs and VFA, respectively. Vertebral fractures (VFs) were 231 in 126 patients and 228 in 125 patients by semiquantitative assessment of radiographs (SQ-Rx) and by VFA, respectively. There was excellent agreement between the two techniques and high diagnostic performance of VFA both on a per-vertebra basis (k score = 0.984, 95 % CI 0.972-0.996, sensitivity 98.68 %, specificity 99.91 %, PPV 98.25 %, NPV 99.93 %) and on a per-patient basis (k score = 0.957, 95 % CI 0.925-0.988, sensitivity 96.83 %, specificity 98.66 %, PPV 97.60 %, NPV 98.22 %). In older patients (≥65 years) affected by moderate or severe osteoarthritis, SQ-Rx and VFA identified 96 VFs and 95 versus 90 vertebral deformities, respectively. This study demonstrates that most vertebrae are evaluable using the iDXA scanner, with improved VFA diagnostic performance even in discriminating mild VFs from vertebral deformities. Therefore, VFA may be appropriate as an alternative to conventional radiography in patients at high risk of VF who are undergoing DXA bone densitometry and in the follow-up of osteoporotic patients on treatment.

Diagnostic Performance of Vertebral Fracture Assessment by the Lunar iDXA Scanner Compared to Conventional Radiography / Diacinti, Daniele; Romano Del, Fiacco; Daniela, Pisani; Federico, Todde; Cattaruzza, Maria Sofia; Diacinti, Davide; Arima, Serena; Elisabetta, Romagnoli; Pepe, Jessica; Cipriani, Cristiana; Minisola, Salvatore. - In: CALCIFIED TISSUE INTERNATIONAL. - ISSN 0171-967X. - STAMPA. - 91:5(2012), pp. 335-342. [10.1007/s00223-012-9643-0]

Diagnostic Performance of Vertebral Fracture Assessment by the Lunar iDXA Scanner Compared to Conventional Radiography

DIACINTI, Daniele;CATTARUZZA, Maria Sofia;DIACINTI, DAVIDE;ARIMA, SERENA;PEPE, JESSICA;CIPRIANI, Cristiana;MINISOLA, Salvatore
2012

Abstract

The purpose of this study was to evaluate the diagnostic performance of vertebral fracture assessment (VFA) using the Lunar iDXA scanner. Conventional spinal radiographs and images acquired by dual-energy X-ray absorptiometry (DXA) of 350 subjects (269 females, 81 males) were evaluated by two different readers. We visualized 4,476/4,550 (98.4 %) vertebrae from T4 to L4 on VFA images compared to 4,535/4,550 (99.7 %) on radiographs. Among the visualized vertebrae, 205/4,535 (4.5 %) and 190/4,476 (4.2 %) were identified as nonfracture deformities by reading of radiographs and VFA, respectively. Vertebral fractures (VFs) were 231 in 126 patients and 228 in 125 patients by semiquantitative assessment of radiographs (SQ-Rx) and by VFA, respectively. There was excellent agreement between the two techniques and high diagnostic performance of VFA both on a per-vertebra basis (k score = 0.984, 95 % CI 0.972-0.996, sensitivity 98.68 %, specificity 99.91 %, PPV 98.25 %, NPV 99.93 %) and on a per-patient basis (k score = 0.957, 95 % CI 0.925-0.988, sensitivity 96.83 %, specificity 98.66 %, PPV 97.60 %, NPV 98.22 %). In older patients (≥65 years) affected by moderate or severe osteoarthritis, SQ-Rx and VFA identified 96 VFs and 95 versus 90 vertebral deformities, respectively. This study demonstrates that most vertebrae are evaluable using the iDXA scanner, with improved VFA diagnostic performance even in discriminating mild VFs from vertebral deformities. Therefore, VFA may be appropriate as an alternative to conventional radiography in patients at high risk of VF who are undergoing DXA bone densitometry and in the follow-up of osteoporotic patients on treatment.
2012
semiquantitative assessment; dual-energy x-ray absorptiometry; vertebral fracture assessment; vertebral fracture; conventional radiography
01 Pubblicazione su rivista::01a Articolo in rivista
Diagnostic Performance of Vertebral Fracture Assessment by the Lunar iDXA Scanner Compared to Conventional Radiography / Diacinti, Daniele; Romano Del, Fiacco; Daniela, Pisani; Federico, Todde; Cattaruzza, Maria Sofia; Diacinti, Davide; Arima, Serena; Elisabetta, Romagnoli; Pepe, Jessica; Cipriani, Cristiana; Minisola, Salvatore. - In: CALCIFIED TISSUE INTERNATIONAL. - ISSN 0171-967X. - STAMPA. - 91:5(2012), pp. 335-342. [10.1007/s00223-012-9643-0]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/483116
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 11
  • Scopus 43
  • ???jsp.display-item.citation.isi??? 41
social impact