In this improved assay only a 500-microL sample is needed and a single assay can be done within 30 min, 10 samples within 90 minutes. The sample of serum or plasma is diluted 20-fold with water, and the estriol is adsorbed from it onto graphitized carbon black ( Carbopack B, Supelco ). After two washings the estriol is desorbed with chloroform/methanol (60/40 by vol), which then is evaporated. The residue is redissolved in 50 microL of water/acetonitrile, and 20 microL is injected into the chromatograph. Analytical recovery for estriol-supplemented serum or plasma averaged 98.6%. Day-to-day CVs ranged from 3.9% at 2 micrograms/L to 2.1% at 20 micrograms/L. The limit of sensitivity is 0.3 micrograms/L, which makes this procedure suitable for determination of estriol even in the first half of pregnancy. Our method is inexpensive, and shows that liquid chromatography can be used to determine estriol in pregnancy serum or plasma. It also is more sensitive and precise and requires less sample than other such methods.

Improved assay of uncon¬juga¬ted estriol in mater¬nal serum or plasma by ad¬sorption and liquid chromatogra¬phy with fluorimetric detection / Andreolini, Franca; Borra, Claudio; DI CORCIA, Antonio; Lagana', Aldo; Samperi, Roberto; Raponi, Giammarco. - In: CLINICAL CHEMISTRY. - ISSN 1056-599X. - STAMPA. - 30:5(1984), pp. 742-744.

Improved assay of uncon¬juga¬ted estriol in mater¬nal serum or plasma by ad¬sorption and liquid chromatogra¬phy with fluorimetric detection

DI CORCIA, Antonio;LAGANA', Aldo;SAMPERI, Roberto;RAPONI, Giammarco
1984

Abstract

In this improved assay only a 500-microL sample is needed and a single assay can be done within 30 min, 10 samples within 90 minutes. The sample of serum or plasma is diluted 20-fold with water, and the estriol is adsorbed from it onto graphitized carbon black ( Carbopack B, Supelco ). After two washings the estriol is desorbed with chloroform/methanol (60/40 by vol), which then is evaporated. The residue is redissolved in 50 microL of water/acetonitrile, and 20 microL is injected into the chromatograph. Analytical recovery for estriol-supplemented serum or plasma averaged 98.6%. Day-to-day CVs ranged from 3.9% at 2 micrograms/L to 2.1% at 20 micrograms/L. The limit of sensitivity is 0.3 micrograms/L, which makes this procedure suitable for determination of estriol even in the first half of pregnancy. Our method is inexpensive, and shows that liquid chromatography can be used to determine estriol in pregnancy serum or plasma. It also is more sensitive and precise and requires less sample than other such methods.
1984
01 Pubblicazione su rivista::01a Articolo in rivista
Improved assay of uncon¬juga¬ted estriol in mater¬nal serum or plasma by ad¬sorption and liquid chromatogra¬phy with fluorimetric detection / Andreolini, Franca; Borra, Claudio; DI CORCIA, Antonio; Lagana', Aldo; Samperi, Roberto; Raponi, Giammarco. - In: CLINICAL CHEMISTRY. - ISSN 1056-599X. - STAMPA. - 30:5(1984), pp. 742-744.
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/474370
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 21
  • ???jsp.display-item.citation.isi??? 21
social impact