A streaming potential in a liq. membrane electrode sensitive to nicotinate ion was obsd. during dynamic electrode response measurements with Na nicotinate stepwise concn. increases. It was possible to sep. the streaming potential from the dynamic electrode response and to study it as a function of the flow-rate in the cell and of the nature of the liq. membrane support. A streaming potential was obsd. with Millipore MF-PH supports (mixed esters of cellulose), but not with Fluoropore (polytetrafluoroethylene bonded to polyethylene net) supports.
Streaming potential in a liquid ion-exchange membrane electrode / Gozzi, Daniele; Gilberto, Scorcelletti. - In: JOURNAL OF ELECTROANALYTICAL CHEMISTRY AND INTERFACIAL ELECTROCHEMISTRY. - ISSN 0022-0728. - STAMPA. - 93:2(1978), pp. 109-115. [10.1016/s0022-0728(78)80223-x]
Streaming potential in a liquid ion-exchange membrane electrode
GOZZI, Daniele;
1978
Abstract
A streaming potential in a liq. membrane electrode sensitive to nicotinate ion was obsd. during dynamic electrode response measurements with Na nicotinate stepwise concn. increases. It was possible to sep. the streaming potential from the dynamic electrode response and to study it as a function of the flow-rate in the cell and of the nature of the liq. membrane support. A streaming potential was obsd. with Millipore MF-PH supports (mixed esters of cellulose), but not with Fluoropore (polytetrafluoroethylene bonded to polyethylene net) supports.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.