The procedure described here shows the advantages of short time, minimal sample, high sensitivity, good accuracy and reproducibility for analyzing VMA concentration in urine. The method is applied to the assay of urine from patients with neuroblastoma, pheochromocytoma and Parkinson's disease. Human urinary VMA values obtained by this method are in agreement with those of other workers
High-performance liquid chromatography procedure for the analysis of urinary 3-methoxy-4-hidroxymandelic acid / Lagana', Aldo; Rotatori, Mauro. - In: JOURNAL OF CHROMATOGRAPHY B. BIOMEDICAL APPLICATIONS. - ISSN 0378-4347. - STAMPA. - 275:1(1983), pp. 168-173. [10.1016/S0378-4347(00)84356-X]
High-performance liquid chromatography procedure for the analysis of urinary 3-methoxy-4-hidroxymandelic acid
LAGANA', Aldo;ROTATORI, MAURO
1983
Abstract
The procedure described here shows the advantages of short time, minimal sample, high sensitivity, good accuracy and reproducibility for analyzing VMA concentration in urine. The method is applied to the assay of urine from patients with neuroblastoma, pheochromocytoma and Parkinson's disease. Human urinary VMA values obtained by this method are in agreement with those of other workersI documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.