Background: Secure tendon-to-bone fixation is essential for successful rotator cuff repair. Biomechanical properties of devices used in rotator cuff repair should be better understood. Purpose: To evaluate the response to incremental cyclic loading of 6 different anchor-suture complexes commonly used in rotator cuff repair. Study Design: Controlled laboratory study. Methods: Two absorbable anchors 5 mm and 6.5 mm in diameter and 1 metallic anchor, coupled with Ethibond or FiberWire, were tested on 5 pairs of fresh-frozen human cadaveric shoulders. An incremental cyclic load was applied until failure using a Zwich-Roell Z010 electromechanical testing machine. The ultimate failure load and mode of failure were recorded. An analysis of variance model was used for statistical analysis. Results: The FiberWire suture coupled with both absorbable and metallic anchors provided statistically significantly stronger fixation. However, although the metallic anchors in most cases failed because of slippage of the anchor, absorbable anchors failed because of rupture of the eyelet. Conclusions: The FiberWire seems to increase the strength of fixation devices under cyclic load using both absorbable and metallic anchors, with relevant differences in failure mode (slippage of the metallic anchor and eyelet failure in the absorbable anchor). Clinical Relevance: Use of the FiberWire suture might change the mode of failure of the suture-anchor complexes.

Effect of cyclic loading on new polyblend suture coupled with different anchors / DE CARLI, Angelo; Ferretti, Andrea; Labianca, Luca; A., Vadala; Monaco, Edoardo; E., Zanzotto. - In: THE AMERICAN JOURNAL OF SPORTS MEDICINE. - ISSN 0363-5465. - STAMPA. - 33:2(2005), pp. 214-219. [10.1177/0363546504267348]

Effect of cyclic loading on new polyblend suture coupled with different anchors

DE CARLI, ANGELO;FERRETTI, Andrea;LABIANCA, LUCA;MONACO, Edoardo;
2005

Abstract

Background: Secure tendon-to-bone fixation is essential for successful rotator cuff repair. Biomechanical properties of devices used in rotator cuff repair should be better understood. Purpose: To evaluate the response to incremental cyclic loading of 6 different anchor-suture complexes commonly used in rotator cuff repair. Study Design: Controlled laboratory study. Methods: Two absorbable anchors 5 mm and 6.5 mm in diameter and 1 metallic anchor, coupled with Ethibond or FiberWire, were tested on 5 pairs of fresh-frozen human cadaveric shoulders. An incremental cyclic load was applied until failure using a Zwich-Roell Z010 electromechanical testing machine. The ultimate failure load and mode of failure were recorded. An analysis of variance model was used for statistical analysis. Results: The FiberWire suture coupled with both absorbable and metallic anchors provided statistically significantly stronger fixation. However, although the metallic anchors in most cases failed because of slippage of the anchor, absorbable anchors failed because of rupture of the eyelet. Conclusions: The FiberWire seems to increase the strength of fixation devices under cyclic load using both absorbable and metallic anchors, with relevant differences in failure mode (slippage of the metallic anchor and eyelet failure in the absorbable anchor). Clinical Relevance: Use of the FiberWire suture might change the mode of failure of the suture-anchor complexes.
2005
anchors; biomechanics; cyclic load; shoulder; sutures
01 Pubblicazione su rivista::01a Articolo in rivista
Effect of cyclic loading on new polyblend suture coupled with different anchors / DE CARLI, Angelo; Ferretti, Andrea; Labianca, Luca; A., Vadala; Monaco, Edoardo; E., Zanzotto. - In: THE AMERICAN JOURNAL OF SPORTS MEDICINE. - ISSN 0363-5465. - STAMPA. - 33:2(2005), pp. 214-219. [10.1177/0363546504267348]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/423893
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 7
  • Scopus 50
  • ???jsp.display-item.citation.isi??? 44
social impact