Recently the application of small imagers in nuclear medicine is growing, particularly in scintimammography. In this paper we propose the use of the Hamamatsu R7600-C8 Position Sensitive Photomultiplier Tube (PSPMT) for detection image optimization in scintimammography through the evaluation of image performances of detector pixellation and its interaction with a collimator lattice . To this aim, a number of CsI(Tl) scintillating arrays with different pixel size and thickness were coupled to the same PSPMT. Considering the very high intrinsic spatial resolution, a look up table was realized to correct accurately the gain non uniformities. The results show an overall energy resolution FWHM, from flood field irradiation source, in close agreement with the individual crystal response and an uniformity response within ± 10%. Finally an analysis of tumor signal to noise ratio (SNR) as a function of detection pixellation was performed, utilizing a breast phantom. Tumor SNR values are highest for 5 mm thick CsI(Tl) arrays but shown slight differences with pixel size. The best tumor SNR value for 6 cm breast thick was obtained with 1.5 × 1.5 mm2 crystal pixel size. The collimator used was a low energy all purpose, parallel hole collimator with 22 mm length, 1.5 mm hexagonal hole, 0.2 mm Pb septa, currently utilized in scintimammography.

Imaging evaluation of CsI(Tl) arrays for scintimammography / Pani, Roberto; Cinti, Maria Nerina; Scafe, R.; Soluri, A.; Pellegrini, Rosanna; DE VINCENTIS, Giuseppe; Trotta, G.; Betti, M.; Burgio, N.; Ciavola, C.; Garibaldi, F.; Cusanno, F.; Zavattini, G.; Di Domenico, G.; Bollini, D.; Lanconelli, N.; Olivo, A.; Pani, S.; Rigon, L.; Del Guerra, A.. - STAMPA. - 3:(2000), pp. 21/105-21/108. (Intervento presentato al convegno 2000 IEEE Nuclear Science Symposium Conference Record tenutosi a Lyon nel 15 October 2000 through 20 October 2000) [10.1109/nssmic.2000.949361].

Imaging evaluation of CsI(Tl) arrays for scintimammography

PANI, Roberto;CINTI, Maria Nerina;PELLEGRINI, Rosanna;DE VINCENTIS, Giuseppe;N. Burgio;
2000

Abstract

Recently the application of small imagers in nuclear medicine is growing, particularly in scintimammography. In this paper we propose the use of the Hamamatsu R7600-C8 Position Sensitive Photomultiplier Tube (PSPMT) for detection image optimization in scintimammography through the evaluation of image performances of detector pixellation and its interaction with a collimator lattice . To this aim, a number of CsI(Tl) scintillating arrays with different pixel size and thickness were coupled to the same PSPMT. Considering the very high intrinsic spatial resolution, a look up table was realized to correct accurately the gain non uniformities. The results show an overall energy resolution FWHM, from flood field irradiation source, in close agreement with the individual crystal response and an uniformity response within ± 10%. Finally an analysis of tumor signal to noise ratio (SNR) as a function of detection pixellation was performed, utilizing a breast phantom. Tumor SNR values are highest for 5 mm thick CsI(Tl) arrays but shown slight differences with pixel size. The best tumor SNR value for 6 cm breast thick was obtained with 1.5 × 1.5 mm2 crystal pixel size. The collimator used was a low energy all purpose, parallel hole collimator with 22 mm length, 1.5 mm hexagonal hole, 0.2 mm Pb septa, currently utilized in scintimammography.
2000
2000 IEEE Nuclear Science Symposium Conference Record
04 Pubblicazione in atti di convegno::04b Atto di convegno in volume
Imaging evaluation of CsI(Tl) arrays for scintimammography / Pani, Roberto; Cinti, Maria Nerina; Scafe, R.; Soluri, A.; Pellegrini, Rosanna; DE VINCENTIS, Giuseppe; Trotta, G.; Betti, M.; Burgio, N.; Ciavola, C.; Garibaldi, F.; Cusanno, F.; Zavattini, G.; Di Domenico, G.; Bollini, D.; Lanconelli, N.; Olivo, A.; Pani, S.; Rigon, L.; Del Guerra, A.. - STAMPA. - 3:(2000), pp. 21/105-21/108. (Intervento presentato al convegno 2000 IEEE Nuclear Science Symposium Conference Record tenutosi a Lyon nel 15 October 2000 through 20 October 2000) [10.1109/nssmic.2000.949361].
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/396656
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 5
  • ???jsp.display-item.citation.isi??? ND
social impact