Our objective was to optimize a scanning protocol for CT colonography using a multidetector scanner with an adaptive-array matrix in a phantom study. A colonic phantom with 27 simulated lesions was examined using a multidetector CT scanner (Somatom Plus 4 Volume Zoom, Siemens, Erlangen, Germany). Three scanning protocols were tested: (a) collimation 1.0 mm, slice thickness 1.0 mm; (b) collimation 2.5 mm, slice thickness 3.0 mm; and (c) collimation 2.5 mm, slice thickness 5.0 mm. Image analysis was performed by two radiologists blinded to the construction of the phantom and to imaging parameters. Sensitivity for lesion detection and image quality was assessed. Computed tomographic colonography detected 27 of 27 lesions with protocol 1 (sensitivity 100%); 26 of 27 lesions with protocol 2 (sensitivity 96.3%); and 23 of 27 lesions with protocol 3 (sensitivity 85.2%). Image quality was graded as optimal for protocol 1, good for protocol 2, and poor for protocol 3. When multidetector CT equipment with an adaptive-array matrix is used, protocols with 1.0- or 3.0-mm slice thickness have a sensitivity for the detection of lesions higher than that obtained with a 5.0-mm slice thickness. Image quality decreases progressively when the slice thickness is increased; therefore, in a clinical setting, we recommend the use of a high-resolution protocol (1.0-mm effective slice thickness) with the highest pitch value in order to reduce scanning time. Dose exposure for such a protocol if combined with 80 mAs is within the range limits recommended by the ICRP for abdominal CT scanning.

Experimental colonic phantom for the evaluation of the optimal scanning technique for CT colonography using a multidetector spiral CT equipment / Laghi, Andrea; R., Iannaccone; F., Mangiapane; F., Piacentini; S., Iori; Passariello, Roberto. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - STAMPA. - 13:(2003), pp. 459-66. [10.1007/s00330-002-1671-0]

Experimental colonic phantom for the evaluation of the optimal scanning technique for CT colonography using a multidetector spiral CT equipment

LAGHI, ANDREA;PASSARIELLO, Roberto
2003

Abstract

Our objective was to optimize a scanning protocol for CT colonography using a multidetector scanner with an adaptive-array matrix in a phantom study. A colonic phantom with 27 simulated lesions was examined using a multidetector CT scanner (Somatom Plus 4 Volume Zoom, Siemens, Erlangen, Germany). Three scanning protocols were tested: (a) collimation 1.0 mm, slice thickness 1.0 mm; (b) collimation 2.5 mm, slice thickness 3.0 mm; and (c) collimation 2.5 mm, slice thickness 5.0 mm. Image analysis was performed by two radiologists blinded to the construction of the phantom and to imaging parameters. Sensitivity for lesion detection and image quality was assessed. Computed tomographic colonography detected 27 of 27 lesions with protocol 1 (sensitivity 100%); 26 of 27 lesions with protocol 2 (sensitivity 96.3%); and 23 of 27 lesions with protocol 3 (sensitivity 85.2%). Image quality was graded as optimal for protocol 1, good for protocol 2, and poor for protocol 3. When multidetector CT equipment with an adaptive-array matrix is used, protocols with 1.0- or 3.0-mm slice thickness have a sensitivity for the detection of lesions higher than that obtained with a 5.0-mm slice thickness. Image quality decreases progressively when the slice thickness is increased; therefore, in a clinical setting, we recommend the use of a high-resolution protocol (1.0-mm effective slice thickness) with the highest pitch value in order to reduce scanning time. Dose exposure for such a protocol if combined with 80 mAs is within the range limits recommended by the ICRP for abdominal CT scanning.
2003
Tomography: Spiral Computed; Sensitivity and Specificity; Phantoms: Imaging; Colonic Polyps; Radiographic Image Interpretation: Computer-Assisted; Colonography: Computed Tomographic; Evaluation Studies as Topic; Models: Theoretical; Humans; Chi-Square Distribution; Probability
01 Pubblicazione su rivista::01a Articolo in rivista
Experimental colonic phantom for the evaluation of the optimal scanning technique for CT colonography using a multidetector spiral CT equipment / Laghi, Andrea; R., Iannaccone; F., Mangiapane; F., Piacentini; S., Iori; Passariello, Roberto. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - STAMPA. - 13:(2003), pp. 459-66. [10.1007/s00330-002-1671-0]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/386979
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 34
  • ???jsp.display-item.citation.isi??? 32
social impact