The barrel part of the CMS electromagnetic calorimeter consists of about 75 000 Lead Tungstate (PbWO4) crystals arranged in 36 x 4 modules which are assembled in two Regional Centres, in Rome and at CERN. Two automatic machines have been designed to check the crystal quality before assembly. The main crystal characteristics are compared to a set of specifications included in the contract with the crystal producers. The measurement stability and cross-calibration between the two machines is a fundamental issue, which has to be monitored throughout the construction phase. This paper describes comparisons between measurements made at the two regional centres to ensure a consistent and reliable crystal quality control. (C) 2004 Elsevier B.V. All rights reserved.

Cross-calibration of two automatic quality control systems for the CMS ECAL crystals / E., Auffray; E., Bouquerel; G., Chevenier; P., Lecoq; L., Perez Prado; P., Sempere Roldan; M., Schneegans; F., Cavallari; S., Costantini; I., Dafinei; M., Diemoz; Longo, Egidio; P., Meridiani; Organtini, Giovanni; Paramatti, Riccardo; S., Baccaro; M., Montecchi. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - 523:3(2004), pp. 355-364. [10.1016/j.nima.2003.11.431]

Cross-calibration of two automatic quality control systems for the CMS ECAL crystals

LONGO, Egidio;ORGANTINI, Giovanni;PARAMATTI, Riccardo;
2004

Abstract

The barrel part of the CMS electromagnetic calorimeter consists of about 75 000 Lead Tungstate (PbWO4) crystals arranged in 36 x 4 modules which are assembled in two Regional Centres, in Rome and at CERN. Two automatic machines have been designed to check the crystal quality before assembly. The main crystal characteristics are compared to a set of specifications included in the contract with the crystal producers. The measurement stability and cross-calibration between the two machines is a fundamental issue, which has to be monitored throughout the construction phase. This paper describes comparisons between measurements made at the two regional centres to ensure a consistent and reliable crystal quality control. (C) 2004 Elsevier B.V. All rights reserved.
2004
calibration; electromagnetic calorimeter; measurement automation; optical properties; scintillating crystals
01 Pubblicazione su rivista::01a Articolo in rivista
Cross-calibration of two automatic quality control systems for the CMS ECAL crystals / E., Auffray; E., Bouquerel; G., Chevenier; P., Lecoq; L., Perez Prado; P., Sempere Roldan; M., Schneegans; F., Cavallari; S., Costantini; I., Dafinei; M., Diemoz; Longo, Egidio; P., Meridiani; Organtini, Giovanni; Paramatti, Riccardo; S., Baccaro; M., Montecchi. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - 523:3(2004), pp. 355-364. [10.1016/j.nima.2003.11.431]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/386093
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 5
social impact