Although rats are commonly used to study left ventricular (LV) hypertrophy, measurement of LV mass and dimensions has required killing the rat. To determine the accuracy of echocardiography in rats, blinded crossectional area (CSA) and LV mass measurements using either the cube function (LVM) or an elliptical model (LVMel) from high resolution M-mode echocardiograms were compared to necropsy LV weight (0.28 to 1.5 g), in 41 normotensive (body weight 116 to 762 g) and 17 hypertensive rats (350 to 560 g). Postmortem chamber volumes in 28 normal rats (0.02 to 0.19 mL) were also compared to echocardiographic volumes derived from the elliptical model. Correlation with LV weight was r = 0.87 for LVM, 0.90 for CSA and 0.93 for LVMel (all P less than .00001). Comparison of hypertensive and body-weight-matched normotensive rats revealed the upper normal limit for both LVMel and CSA to have 89% sensitivity and 100% specificity for detection of post mortem LV hypertrophy. Necropsy LV volumes were more closely related to systolic echocardiographic volumes than to diastolic volumes (r = 0.78 v 0.71, both P less than .00001), compatible with the effects of post mortem contracture. Stroke volume determined invasively in 5 Wistar rats by thermodilution was similar to that obtained using elliptical model echo volumes in 5 rats of the same body size (0.35 +/- 0.05 v 0.30 +/- 0.06 mL/beat). Echocardiography can be used to evaluate LV structure and function in rats and to detect in vivo LV anatomic differences induced by hypertension.

Echocardiographic measurement of left ventricular mass and volume in normotensive and hypertensive rats. Necropsy validation / de Simone, G; Wallerson, Dc; Volpe, Massimo; Devereux, Rb. - In: AMERICAN JOURNAL OF HYPERTENSION. - ISSN 0895-7061. - 3:(1990), pp. 688-696.

Echocardiographic measurement of left ventricular mass and volume in normotensive and hypertensive rats. Necropsy validation.

VOLPE, Massimo;
1990

Abstract

Although rats are commonly used to study left ventricular (LV) hypertrophy, measurement of LV mass and dimensions has required killing the rat. To determine the accuracy of echocardiography in rats, blinded crossectional area (CSA) and LV mass measurements using either the cube function (LVM) or an elliptical model (LVMel) from high resolution M-mode echocardiograms were compared to necropsy LV weight (0.28 to 1.5 g), in 41 normotensive (body weight 116 to 762 g) and 17 hypertensive rats (350 to 560 g). Postmortem chamber volumes in 28 normal rats (0.02 to 0.19 mL) were also compared to echocardiographic volumes derived from the elliptical model. Correlation with LV weight was r = 0.87 for LVM, 0.90 for CSA and 0.93 for LVMel (all P less than .00001). Comparison of hypertensive and body-weight-matched normotensive rats revealed the upper normal limit for both LVMel and CSA to have 89% sensitivity and 100% specificity for detection of post mortem LV hypertrophy. Necropsy LV volumes were more closely related to systolic echocardiographic volumes than to diastolic volumes (r = 0.78 v 0.71, both P less than .00001), compatible with the effects of post mortem contracture. Stroke volume determined invasively in 5 Wistar rats by thermodilution was similar to that obtained using elliptical model echo volumes in 5 rats of the same body size (0.35 +/- 0.05 v 0.30 +/- 0.06 mL/beat). Echocardiography can be used to evaluate LV structure and function in rats and to detect in vivo LV anatomic differences induced by hypertension.
1990
01 Pubblicazione su rivista::01a Articolo in rivista
Echocardiographic measurement of left ventricular mass and volume in normotensive and hypertensive rats. Necropsy validation / de Simone, G; Wallerson, Dc; Volpe, Massimo; Devereux, Rb. - In: AMERICAN JOURNAL OF HYPERTENSION. - ISSN 0895-7061. - 3:(1990), pp. 688-696.
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/382194
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 17
  • Scopus 120
  • ???jsp.display-item.citation.isi??? 113
social impact