A first prototype of optical scanner based on a CCD camera and the use of Fricke-agarose-XO has allowed us to reconstruct 3D dose maps, monitoring doses in the range 0.5-3 Gy. Measurements performed with laboratory tests and with clinical beams would indicate that the uncertainty involved in dose mapping are compatible with requirements of a 3D dosimetry for the verifications of treatment plans. © 2009 IOP Publishing Ltd.
Dosimetric performances of optically detected Fricke-agarose-Xylenol Orange gel / A. M., Luciani; A., Palma; S., Grande; P., Sordi; L., Guidoni; V., Viti; DI CASTRO, Elisabetta; C., Di Felice; A., Lo Bosco. - In: JOURNAL OF PHYSICS. CONFERENCE SERIES. - ISSN 1742-6588. - STAMPA. - 164:(2009). ( 5th International Conference on Radiotherapy Gel Dosimetry (DOSGEL 2008) Hersonissos; Crete ) [10.1088/1742-6596/164/1/012015].
Dosimetric performances of optically detected Fricke-agarose-Xylenol Orange gel
DI CASTRO, Elisabetta;
2009
Abstract
A first prototype of optical scanner based on a CCD camera and the use of Fricke-agarose-XO has allowed us to reconstruct 3D dose maps, monitoring doses in the range 0.5-3 Gy. Measurements performed with laboratory tests and with clinical beams would indicate that the uncertainty involved in dose mapping are compatible with requirements of a 3D dosimetry for the verifications of treatment plans. © 2009 IOP Publishing Ltd.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.