We are developing a positron emission mammograph imaging system that consists of 2 planar heads (10 x 10 cm 2), each composed of 4 matrices (5 x 5 cm(2)) of scintillating crystals coupled to the same number of multi-anode PMT (Hamamatsu H8500). We are developing a readout scheme which handles the 4 x 64 output signals of each head. We are primarily interested in charge multiplexed readout schemes. The measurements presented in this paper highlight the limits of a multiplexed readout using a discretized positioning circuit (DPC) by comparison with results using either another multiplexed scheme based on symmetric charge division (SCD) or a multi-anode readout. The measurements have been performed with the H8500 PMT coupled to pixelated NaI(Tl) and YAP crystals. (c) 2006 Elsevier B.V. All rights reserved.

Preliminary study of different readout strategies for a positron emission mammograph head / S., Vecchio; M., Camarda; N., Belcari; D., Bianchi; M. N., Cinti; A., Del Guerra; D., Herbert; M., Lazzarotti; Pani, Roberto. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - STAMPA. - 569:2 SPEC. ISS.(2006), pp. 264-268. (Intervento presentato al convegno 3rd International Conference on Imaging Technologies in Biomedical Sciences tenutosi a Milos, GREECE nel SEP 25-29, 2005) [10.1016/j.nima.2006.08.025].

Preliminary study of different readout strategies for a positron emission mammograph head

PANI, Roberto
2006

Abstract

We are developing a positron emission mammograph imaging system that consists of 2 planar heads (10 x 10 cm 2), each composed of 4 matrices (5 x 5 cm(2)) of scintillating crystals coupled to the same number of multi-anode PMT (Hamamatsu H8500). We are developing a readout scheme which handles the 4 x 64 output signals of each head. We are primarily interested in charge multiplexed readout schemes. The measurements presented in this paper highlight the limits of a multiplexed readout using a discretized positioning circuit (DPC) by comparison with results using either another multiplexed scheme based on symmetric charge division (SCD) or a multi-anode readout. The measurements have been performed with the H8500 PMT coupled to pixelated NaI(Tl) and YAP crystals. (c) 2006 Elsevier B.V. All rights reserved.
2006
dpc; flat panel; multi-anode readout; multiplexed readout; position-sensitive photomultiplier; scd
01 Pubblicazione su rivista::01a Articolo in rivista
Preliminary study of different readout strategies for a positron emission mammograph head / S., Vecchio; M., Camarda; N., Belcari; D., Bianchi; M. N., Cinti; A., Del Guerra; D., Herbert; M., Lazzarotti; Pani, Roberto. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - STAMPA. - 569:2 SPEC. ISS.(2006), pp. 264-268. (Intervento presentato al convegno 3rd International Conference on Imaging Technologies in Biomedical Sciences tenutosi a Milos, GREECE nel SEP 25-29, 2005) [10.1016/j.nima.2006.08.025].
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/365354
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 9
social impact