The evaluation of mechanical behavior of plastic Ankle-Foot Orthosis (AFO) is important since AFO can provide an efficient support to patients with disabilities in locomotion. This paper reports on a novel testing apparatus that allows: (a) the evaluation of AFO stiffness in sagittal and frontal planes; (b) the conduction of semi-automatic trials; and, finally, (c) a global accuracy associated to the AFO stiffness values always less than 4%. The stiffness values are determined by the measurements of the imposed relative displacements between the foot and the shank of the orthosis and the induced reaction forces. The data collected together in an exact 2-D approach, together with those provided by gait analysis systems, allows to better understand gait alteration induced by ankle orthosis, and to improve clinical management of patients.

A Novel Device to Evaluate the Stiffness of Ankle-Foot Orthosis Devices / Cappa, Paolo; Patane', Fabrizio; M. M., Pierro. - In: JOURNAL OF BIOMECHANICAL ENGINEERING. - ISSN 0148-0731. - STAMPA. - 125:6(2003), pp. 913-917. [10.1115/1.1634993]

A Novel Device to Evaluate the Stiffness of Ankle-Foot Orthosis Devices

CAPPA, Paolo;PATANE', FABRIZIO;
2003

Abstract

The evaluation of mechanical behavior of plastic Ankle-Foot Orthosis (AFO) is important since AFO can provide an efficient support to patients with disabilities in locomotion. This paper reports on a novel testing apparatus that allows: (a) the evaluation of AFO stiffness in sagittal and frontal planes; (b) the conduction of semi-automatic trials; and, finally, (c) a global accuracy associated to the AFO stiffness values always less than 4%. The stiffness values are determined by the measurements of the imposed relative displacements between the foot and the shank of the orthosis and the induced reaction forces. The data collected together in an exact 2-D approach, together with those provided by gait analysis systems, allows to better understand gait alteration induced by ankle orthosis, and to improve clinical management of patients.
2003
01 Pubblicazione su rivista::01a Articolo in rivista
A Novel Device to Evaluate the Stiffness of Ankle-Foot Orthosis Devices / Cappa, Paolo; Patane', Fabrizio; M. M., Pierro. - In: JOURNAL OF BIOMECHANICAL ENGINEERING. - ISSN 0148-0731. - STAMPA. - 125:6(2003), pp. 913-917. [10.1115/1.1634993]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/360563
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 12
social impact