We have developed a tomograph for single photon emission imaging (SPECT) of the breast for the detection of small size tumors. The SPECT is mounted on a ring that is rotating around the breast with the patient in prone position. The breast will be imaged by two opposing detector heads of approximately 5 x 15 cm(2) each, with a field of view about 13 cm wide. Each head is made up of one pixilated NaI crystal matrix coupled to three Hamamatsu H8500 PMTs. A "general purpose" lead collimator is positioned in front of the crystal. Detailed simulations have been made for the optimization and the evaluation of the detector performances. Monte Carlo results indicate that tumors of 8 mm diameter are detectable with a tumor/background uptake ratio of 5:1. The experimental characterization of the detector head is presented. The rotating ring is now being assembled. (C) 2007 Elsevier B.V. All rights reserved.

A single photon emission computer tomograph for breast cancer imaging / S., Vecchio; N., Belcari; P., Bennati; M., Camarda; R., Campanini; Cinti, Maria Nerina; A., Del Guerra; E., Iampieri; N., Lanconelli; Pani, Roberto; L., Spontoni. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - STAMPA. - 581:1-2 SPEC. ISS.(2007), pp. 84-87. (Intervento presentato al convegno 11th International Vienna Conference on Instrumentation tenutosi a Vienna, AUSTRIA nel FEB 19-24, 2007) [10.1016/j.nima.2007.07.034].

A single photon emission computer tomograph for breast cancer imaging

CINTI, Maria Nerina;PANI, Roberto;
2007

Abstract

We have developed a tomograph for single photon emission imaging (SPECT) of the breast for the detection of small size tumors. The SPECT is mounted on a ring that is rotating around the breast with the patient in prone position. The breast will be imaged by two opposing detector heads of approximately 5 x 15 cm(2) each, with a field of view about 13 cm wide. Each head is made up of one pixilated NaI crystal matrix coupled to three Hamamatsu H8500 PMTs. A "general purpose" lead collimator is positioned in front of the crystal. Detailed simulations have been made for the optimization and the evaluation of the detector performances. Monte Carlo results indicate that tumors of 8 mm diameter are detectable with a tumor/background uptake ratio of 5:1. The experimental characterization of the detector head is presented. The rotating ring is now being assembled. (C) 2007 Elsevier B.V. All rights reserved.
2007
breast cancer; ma-pmt; spect
01 Pubblicazione su rivista::01a Articolo in rivista
A single photon emission computer tomograph for breast cancer imaging / S., Vecchio; N., Belcari; P., Bennati; M., Camarda; R., Campanini; Cinti, Maria Nerina; A., Del Guerra; E., Iampieri; N., Lanconelli; Pani, Roberto; L., Spontoni. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - STAMPA. - 581:1-2 SPEC. ISS.(2007), pp. 84-87. (Intervento presentato al convegno 11th International Vienna Conference on Instrumentation tenutosi a Vienna, AUSTRIA nel FEB 19-24, 2007) [10.1016/j.nima.2007.07.034].
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/360288
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 8
social impact