In this paper the interaction between the radiofrequency field produced by magnetic resonance imaging apparatus and a pacemaker (PM) implanted inside an accurate anatomical model of the human body is studied. The interaction of the PM catheter with the EM field gives rise to strong local SAR values inside the heart where the catheter tip is fixed. However the SAR values averaged over 10 g remain below the limits reported in the ICNIRP recommendations.
Interaction between pacemakers implanted in realistic human models and the radio frequency field produced by magnetic resonance imaging apparatus / Pisa, Stefano; G., Calcagnini; Cavagnaro, Marta; Piuzzi, Emanuele; E., Mattei; Bernardi, Paolo Italo. - (2007), pp. 982-985. ( 2007 ACES - 23rd Annual Review of Progress in Applied Computational Electromagnetics Verona ).
Interaction between pacemakers implanted in realistic human models and the radio frequency field produced by magnetic resonance imaging apparatus
PISA, Stefano;CAVAGNARO, Marta;PIUZZI, Emanuele;BERNARDI, Paolo Italo
2007
Abstract
In this paper the interaction between the radiofrequency field produced by magnetic resonance imaging apparatus and a pacemaker (PM) implanted inside an accurate anatomical model of the human body is studied. The interaction of the PM catheter with the EM field gives rise to strong local SAR values inside the heart where the catheter tip is fixed. However the SAR values averaged over 10 g remain below the limits reported in the ICNIRP recommendations.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.