The performance of the new prototype of high quantum efficiency PMT (43% at 380 nm), Hamamatsu R7600U-200, was studied coupled to a LaBr3:Ce crystal with the size of ∅12.5 mm×12.5 mm. The energy resolution results were compared with ones from two PMTs, Hamamatsu R7600U and R6231MOD, with 22% and 30% quantum efficiency (QE), respectively. Moreover, the photodetectors were equipped with tapered and un-tapered voltage dividers to study the non-linearity effects on pulse height distribution, due to very high peak currents induced in the PMT by the fast and intense light pulse of LaBr3:Ce. The results show an energy resolution improvement with UBA PMT of about 20%, in the energy range of 80-662 keV, with respect to the BA one. © 2009 Elsevier B.V. All rights reserved.

Energy resolution measurements of LaBr3:Ce scintillating crystals with an ultra-high quantum efficiency photomultiplier tube / Pani, Roberto; M. N., Cinti; R., Scafe; Pellegrini, Rosanna; F., Vittorini; P., Bennati; S., Ridolfi; S., Lo Meo; Mattioli, Mario Claudio; G., Baldazzi; F., Pisacane; F., Navarria; G., Moschini; P., Boccaccio; V. O., Cencelli; D., Sacco. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - STAMPA. - 610:1(2009), pp. 41-44. [10.1016/j.nima.2009.05.058]

Energy resolution measurements of LaBr3:Ce scintillating crystals with an ultra-high quantum efficiency photomultiplier tube

PANI, Roberto;PELLEGRINI, Rosanna;MATTIOLI, Mario Claudio;
2009

Abstract

The performance of the new prototype of high quantum efficiency PMT (43% at 380 nm), Hamamatsu R7600U-200, was studied coupled to a LaBr3:Ce crystal with the size of ∅12.5 mm×12.5 mm. The energy resolution results were compared with ones from two PMTs, Hamamatsu R7600U and R6231MOD, with 22% and 30% quantum efficiency (QE), respectively. Moreover, the photodetectors were equipped with tapered and un-tapered voltage dividers to study the non-linearity effects on pulse height distribution, due to very high peak currents induced in the PMT by the fast and intense light pulse of LaBr3:Ce. The results show an energy resolution improvement with UBA PMT of about 20%, in the energy range of 80-662 keV, with respect to the BA one. © 2009 Elsevier B.V. All rights reserved.
2009
photodetectors; photomultipliers; scintillation detectors
01 Pubblicazione su rivista::01a Articolo in rivista
Energy resolution measurements of LaBr3:Ce scintillating crystals with an ultra-high quantum efficiency photomultiplier tube / Pani, Roberto; M. N., Cinti; R., Scafe; Pellegrini, Rosanna; F., Vittorini; P., Bennati; S., Ridolfi; S., Lo Meo; Mattioli, Mario Claudio; G., Baldazzi; F., Pisacane; F., Navarria; G., Moschini; P., Boccaccio; V. O., Cencelli; D., Sacco. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - STAMPA. - 610:1(2009), pp. 41-44. [10.1016/j.nima.2009.05.058]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/227595
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 7
social impact