To compare intraoperative efficiency and early postoperative corneal endothelial parameters between a balanced metal phacoemulsification tip and a hybrid polymer-coated tip. In this prospective randomized clinical study, patients with grade 3 nuclear cataract were assigned to phacoemulsification using either a balanced metal tip or a hybrid polymer-coated tip. The primary outcome was total ultrasound time, assessed within a noninferiority framework. Secondary outcomes included cumulative dissipated energy (CDE), average torsional amplitude, aspiration time, fluid usage, intraoperative complications, and endothelial parameters at postoperative day 7. A total of 470 eyes were analyzed (balanced, n = 236; hybrid, n = 234). The hybrid tip met the predefined noninferiority criterion for total ultrasound time (mean difference, hybrid minus balanced, − 10.18 s; 95% CI − 15.28 to − 5.09) and showed a shorter observed ultrasound time than the balanced tip (median, 43.9 vs. 58.8 s; p < 0.001). CDE and torsional amplitude were higher, whereas aspiration time and fluid usage were lower with the hybrid tip (all p < 0.001). Postoperative endothelial data were available for 375 eyes. At day 7, the hybrid group showed higher endothelial cell density, lower endothelial cell loss, and fewer eyes with > 10% endothelial loss. In eyes with grade 3 nuclear cataract, the hybrid polymer-coated tip was noninferior to the balanced metal tip for total ultrasound time. Although CDE was higher, the hybrid tip was associated with shorter ultrasound exposure, lower aspiration time and fluid usage, and a more favorable early endothelial profile.

Balanced versus hybrid phacoemulsification tips in grade 3 nuclear cataracts: a prospective randomized clinical study / Visioli, G., Gagliardi, O.M., Ciccare, C., Iannetti, L., Albanese, G.M., Romaniello, A., Alisi, L., Marenco, M., Lambiase, A., Gharbiya, M.. - In: SCIENTIFIC REPORTS. - ISSN 2045-2322. - (2026). [10.1038/s41598-026-73160-z]

Balanced versus hybrid phacoemulsification tips in grade 3 nuclear cataracts: a prospective randomized clinical study

Giacomo Visioli
Primo
Writing – Original Draft Preparation
;
Oscar Matteo Gagliardi
Secondo
Investigation
;
Chiara Ciccare
Investigation
;
Ludovico Iannetti
Writing – Review & Editing
;
Giuseppe Maria Albanese
Resources
;
Annalisa Romaniello
Investigation
;
Ludovico Alisi
Methodology
;
Marco Marenco
Writing – Review & Editing
;
Alessandro Lambiase
Penultimo
Supervision
;
Magda Gharbiya
Ultimo
Conceptualization
2026

Abstract

To compare intraoperative efficiency and early postoperative corneal endothelial parameters between a balanced metal phacoemulsification tip and a hybrid polymer-coated tip. In this prospective randomized clinical study, patients with grade 3 nuclear cataract were assigned to phacoemulsification using either a balanced metal tip or a hybrid polymer-coated tip. The primary outcome was total ultrasound time, assessed within a noninferiority framework. Secondary outcomes included cumulative dissipated energy (CDE), average torsional amplitude, aspiration time, fluid usage, intraoperative complications, and endothelial parameters at postoperative day 7. A total of 470 eyes were analyzed (balanced, n = 236; hybrid, n = 234). The hybrid tip met the predefined noninferiority criterion for total ultrasound time (mean difference, hybrid minus balanced, − 10.18 s; 95% CI − 15.28 to − 5.09) and showed a shorter observed ultrasound time than the balanced tip (median, 43.9 vs. 58.8 s; p < 0.001). CDE and torsional amplitude were higher, whereas aspiration time and fluid usage were lower with the hybrid tip (all p < 0.001). Postoperative endothelial data were available for 375 eyes. At day 7, the hybrid group showed higher endothelial cell density, lower endothelial cell loss, and fewer eyes with > 10% endothelial loss. In eyes with grade 3 nuclear cataract, the hybrid polymer-coated tip was noninferior to the balanced metal tip for total ultrasound time. Although CDE was higher, the hybrid tip was associated with shorter ultrasound exposure, lower aspiration time and fluid usage, and a more favorable early endothelial profile.
2026
Phacoemulsification;cataract surgery; Phaco tip; Hybrid tip; Cumulative dissipated energy; Ultrasound time; Endothelial cell loss
01 Pubblicazione su rivista::01a Articolo in rivista
Balanced versus hybrid phacoemulsification tips in grade 3 nuclear cataracts: a prospective randomized clinical study / Visioli, G., Gagliardi, O.M., Ciccare, C., Iannetti, L., Albanese, G.M., Romaniello, A., Alisi, L., Marenco, M., Lambiase, A., Gharbiya, M.. - In: SCIENTIFIC REPORTS. - ISSN 2045-2322. - (2026). [10.1038/s41598-026-73160-z]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1781630
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact