Oncological radiative hyperthermia (HT) is a therapy aimed at heating the tumour using electromagnetic (EM) fields. The synergic application of HT and radio/chemotherapy improves the outcome of these cancer therapies. Superficial HT is applied to tumours up to 3-4 cm depth. Typical working frequencies are 434 MHz and 915 MHz. Quality assurance (QA) protocols from the European Society for Hyperthermic Oncology (ESHO) recommend using multilayer tissue-equivalent phantoms to validate HT devices. The current guidelines propose the evaluation of thermal dose parameters for QA procedures. This work presents a numerical study simulating the application of a commercial superficial HT antenna on the multilayer phantom, investigating the relation between electromagnetic and thermal dose parameters. The results aim to support and refine QA procedures for superficial hyperthermia treatments, in order to consider the feasibility of both EM and thermal measurements for the validation of HT devices.

A Numerical Analysis of the Relation between Electromagnetic and Thermal Dose Parameters in Quality Assurance Procedures for Superficial Hyperthermia / Di Cristofano, M., Cavagnaro, M.. - (2026), pp. 1-5. (20th European Conference on Antennas and Propagation, EuCAP 2026 Dublin, Ireland ) [10.23919/EuCAP68105.2026.11612122].

A Numerical Analysis of the Relation between Electromagnetic and Thermal Dose Parameters in Quality Assurance Procedures for Superficial Hyperthermia

Di Cristofano M.;Cavagnaro M.
2026

Abstract

Oncological radiative hyperthermia (HT) is a therapy aimed at heating the tumour using electromagnetic (EM) fields. The synergic application of HT and radio/chemotherapy improves the outcome of these cancer therapies. Superficial HT is applied to tumours up to 3-4 cm depth. Typical working frequencies are 434 MHz and 915 MHz. Quality assurance (QA) protocols from the European Society for Hyperthermic Oncology (ESHO) recommend using multilayer tissue-equivalent phantoms to validate HT devices. The current guidelines propose the evaluation of thermal dose parameters for QA procedures. This work presents a numerical study simulating the application of a commercial superficial HT antenna on the multilayer phantom, investigating the relation between electromagnetic and thermal dose parameters. The results aim to support and refine QA procedures for superficial hyperthermia treatments, in order to consider the feasibility of both EM and thermal measurements for the validation of HT devices.
2026
20th European Conference on Antennas and Propagation, EuCAP 2026
electromagnetic fields; electromagnetic-thermal phenomenon; QA procedures; radiative hyperthermia
04 Pubblicazione in atti di convegno::04b Atto di convegno in volume
A Numerical Analysis of the Relation between Electromagnetic and Thermal Dose Parameters in Quality Assurance Procedures for Superficial Hyperthermia / Di Cristofano, M., Cavagnaro, M.. - (2026), pp. 1-5. (20th European Conference on Antennas and Propagation, EuCAP 2026 Dublin, Ireland ) [10.23919/EuCAP68105.2026.11612122].
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1776469
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact