Introduction: Intradural extramedullary and intramedullary spinal tumors are rare, complex to treat, and require advanced surgical techniques. Ultrasonic aspirators, commonly used for tumor removal, can cause sensory and motor deficits, including loss of motor evoked potentials (MEPs). This study aims to evaluate the safety and efficacy of ultrasonic aspirators in intramedullary tumor surgery using a swine model, comparing different systems and techniques. Methods: Ten pigs underwent D1-D3 laminectomy and myelotomy, with adipose tissue simulating a tumor. The ultrasonic aspirators were tested under varying conditions (fragmentation power, suction, application time, and vibration mode). The primary endpoint is to evaluate the impact of the chosen variables on motor function damage. The secondary endpoints are histological evaluation of the type of damage caused by ultrasound aspirators and the effect of steroid drugs on MEPs’ impairment recovery. Results: Ultrasound aspirators can cause a significant MEP signal reduction when used in continuous mode, with fragmentation power >30 for more than 2 min (p < 0.001). Suction does not affect MEPs. When used in alternating/pulsatile mode, fragmentation power and application time do not affect MEPs. The two-way ANOVA analysis on the interaction between fragmentation power and application time in continuous mode did not demonstrate a significant interaction (p = 0.155). Time alone does not affect motor damage (p = 0.873). Betamethasone can restore MEPs’ signal after damage if administered immediately. Conclusions: Using ultrasonic aspirators in an animal model of intramedullary tumor surgery is safe. The main factor that resulted in the responsibility of motor function impairment is the fragmentation power.
Safety and Efficacy Evaluation of Ultrasound Aspirators in Intramedullary Spinal Cord Tumors Surgery: An Experimental Study on a Swine Model / Palmieri, Mauro; Pesce, Alessandro; Capobianco, Mattia; Corsini, Massimo; Iovannitti, Giorgia; Aloj, Fulvio; Zancana, Giuseppa; Esposito, Vincenzo; Salvati, Maurizio; Santoro, Antonio; Cantore, Gianpaolo; Frati, Alessandro. - In: BRAIN SCIENCES. - ISSN 2076-3425. - 15:7(2025), pp. 1-14. [10.3390/brainsci15070670]
Safety and Efficacy Evaluation of Ultrasound Aspirators in Intramedullary Spinal Cord Tumors Surgery: An Experimental Study on a Swine Model
Mauro Palmieri;Mattia Capobianco;Massimo Corsini;Giorgia Iovannitti;Giuseppa Zancana;Vincenzo Esposito;Antonio Santoro;Alessandro Frati
2025
Abstract
Introduction: Intradural extramedullary and intramedullary spinal tumors are rare, complex to treat, and require advanced surgical techniques. Ultrasonic aspirators, commonly used for tumor removal, can cause sensory and motor deficits, including loss of motor evoked potentials (MEPs). This study aims to evaluate the safety and efficacy of ultrasonic aspirators in intramedullary tumor surgery using a swine model, comparing different systems and techniques. Methods: Ten pigs underwent D1-D3 laminectomy and myelotomy, with adipose tissue simulating a tumor. The ultrasonic aspirators were tested under varying conditions (fragmentation power, suction, application time, and vibration mode). The primary endpoint is to evaluate the impact of the chosen variables on motor function damage. The secondary endpoints are histological evaluation of the type of damage caused by ultrasound aspirators and the effect of steroid drugs on MEPs’ impairment recovery. Results: Ultrasound aspirators can cause a significant MEP signal reduction when used in continuous mode, with fragmentation power >30 for more than 2 min (p < 0.001). Suction does not affect MEPs. When used in alternating/pulsatile mode, fragmentation power and application time do not affect MEPs. The two-way ANOVA analysis on the interaction between fragmentation power and application time in continuous mode did not demonstrate a significant interaction (p = 0.155). Time alone does not affect motor damage (p = 0.873). Betamethasone can restore MEPs’ signal after damage if administered immediately. Conclusions: Using ultrasonic aspirators in an animal model of intramedullary tumor surgery is safe. The main factor that resulted in the responsibility of motor function impairment is the fragmentation power.| File | Dimensione | Formato | |
|---|---|---|---|
|
brainsci-15-00670-v2.pdf
accesso aperto
Note: Palmieri_Safety_2025
Tipologia:
Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza:
Creative commons
Dimensione
2.33 MB
Formato
Adobe PDF
|
2.33 MB | Adobe PDF |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.