The Cleopatra ASIC is a 12-channel prototype ASIC for the readout of hydrogenated amorphous silicon sensors used for real-time dosimetry in radiation diagnostic and radiation therapy. The architecture is based on a current to frequency conversion based on the recycling integrator principle in order to cover a dynamic range of four orders of magnitude with high linearity. Three different input amplifier configurations have been implemented in order to check the trade-off between detector capacitance and maximum output frequency. Cleopatra has been designed in CMOS 28 nm technology and succesfully tested in laboratory.

Cleopatra: a 12-channel recycling integrator ASIC for the readout of hydrogenated amorphous silicon detectors in radiotherapy dosimetry / Mazza, G., Wheadon, R., Lenta, F., Buti, C., Fabi, M., Grimani, C., Pallotta, S., Sabbatini, F., Talamonti, C., Villani, M., Aziz, S., Calcagnile, L., Caricato, A.P., Martino, M., Maruccio, G., Monteduro, A.G., Quarta, G., Rizzato, S., Catalano, R., Cirrone, G.A.P., et al.. - In: JOURNAL OF INSTRUMENTATION. - ISSN 1748-0221. - 20:1(2025). [10.1088/1748-0221/20/01/C01034]

Cleopatra: a 12-channel recycling integrator ASIC for the readout of hydrogenated amorphous silicon detectors in radiotherapy dosimetry

Caputo D.;De Cesare G.;Lovecchio N.;Nascetti A.;
2025

Abstract

The Cleopatra ASIC is a 12-channel prototype ASIC for the readout of hydrogenated amorphous silicon sensors used for real-time dosimetry in radiation diagnostic and radiation therapy. The architecture is based on a current to frequency conversion based on the recycling integrator principle in order to cover a dynamic range of four orders of magnitude with high linearity. Three different input amplifier configurations have been implemented in order to check the trade-off between detector capacitance and maximum output frequency. Cleopatra has been designed in CMOS 28 nm technology and succesfully tested in laboratory.
2025
analogue electronic circuits; digital electronic circuits; dosimetry concepts and apparatus; front-end electronics for detector readout
01 Pubblicazione su rivista::01a Articolo in rivista
Cleopatra: a 12-channel recycling integrator ASIC for the readout of hydrogenated amorphous silicon detectors in radiotherapy dosimetry / Mazza, G., Wheadon, R., Lenta, F., Buti, C., Fabi, M., Grimani, C., Pallotta, S., Sabbatini, F., Talamonti, C., Villani, M., Aziz, S., Calcagnile, L., Caricato, A.P., Martino, M., Maruccio, G., Monteduro, A.G., Quarta, G., Rizzato, S., Catalano, R., Cirrone, G.A.P., et al.. - In: JOURNAL OF INSTRUMENTATION. - ISSN 1748-0221. - 20:1(2025). [10.1088/1748-0221/20/01/C01034]
File allegati a questo prodotto
File Dimensione Formato  
Mazza_Cleopatra_2025.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 1.23 MB
Formato Adobe PDF
1.23 MB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1735649
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact