Background: The subcutaneous implantable cardioverter-defibrillator (S-ICD) is an alternative to traditional ICDs. The PRAETORIAN score, based on chest radiographs, has been validated to predict the probability of successful S-ICD defibrillation testing by assessing factors like fat thickness between the coil and sternum and generator placement. Objective: This study evaluated the correlation between the PRAETORIAN score and clinical characteristics, as well as implantation variables. Methods: We retrospectively analyzed data from 1253 patients who had undergone implantation of an S-ICD across 33 centers. The intermuscular positioning of the pulse generator was adopted in all patients. Post-implantation posterior-anterior and lateral chest radiographs were analyzed to calculate the PRAETORIAN score. Results: 95.7% of patients had a PRAETORIAN score <90, indicative of a low risk of conversion failure. Body mass index (BMI) was the only independent predictor of a score ≥90, and all patients with BMI <25kg/m2 (normal weight or underweight) had a score <90. The intermuscular positioning technique resulted in optimal posterior placement of the device in all patients and significant sub-generator fat in only 3% of cases. A shock impedance value >88Ohm enabled to detect a PRAETORIAN score ≥90 with 98% (95% CI 97%-99%) negative predictive value. Conclusions: In contemporary practice, the PRAETORIAN score can be simplified. By adopting an intermuscular approach, two of the three steps of the score-evaluating the adequate posterior positioning of the generator and measuring the sub-generator fat-become superfluous, and impedance may serve as a reliable surrogate of sub-coil fat thickness. Furthermore, our data suggest that for non-obese patients, a favorable PRAETORIAN score is assured, making the score evaluation potentially unnecessary.

The risk of failure of subcutaneous implantable cardioverter defibrillator therapy: from PRAETORIAN score to clinical practice / Ziacchi, Matteo; Ottaviano, Luca; Checchi, Luca; Viani, Stefano; Nigro, Gerardo; Bianchi, Valter; De Bonis, Silvana; De Filippo, Paolo; Francia, Pietro; Rapacciuolo, Antonio; Vitulano, Gennaro; Perego, Giovanni Battista; Schillaci, Vincenzo; Lavalle, Carlo; Migliore, Federico; Pisanò, Ennio C L; Compagnucci, Paolo; Palmisano, Pietro; Botto, Gianluca; Rordorf, Roberto; Lovecchio, Mariolina; Valsecchi, Sergio; Biffi, Mauro. - In: EUROPACE. - ISSN 1099-5129. - 27:2(2025). [10.1093/europace/euaf011]

The risk of failure of subcutaneous implantable cardioverter defibrillator therapy: from PRAETORIAN score to clinical practice

Francia, Pietro;Lavalle, Carlo;
2025

Abstract

Background: The subcutaneous implantable cardioverter-defibrillator (S-ICD) is an alternative to traditional ICDs. The PRAETORIAN score, based on chest radiographs, has been validated to predict the probability of successful S-ICD defibrillation testing by assessing factors like fat thickness between the coil and sternum and generator placement. Objective: This study evaluated the correlation between the PRAETORIAN score and clinical characteristics, as well as implantation variables. Methods: We retrospectively analyzed data from 1253 patients who had undergone implantation of an S-ICD across 33 centers. The intermuscular positioning of the pulse generator was adopted in all patients. Post-implantation posterior-anterior and lateral chest radiographs were analyzed to calculate the PRAETORIAN score. Results: 95.7% of patients had a PRAETORIAN score <90, indicative of a low risk of conversion failure. Body mass index (BMI) was the only independent predictor of a score ≥90, and all patients with BMI <25kg/m2 (normal weight or underweight) had a score <90. The intermuscular positioning technique resulted in optimal posterior placement of the device in all patients and significant sub-generator fat in only 3% of cases. A shock impedance value >88Ohm enabled to detect a PRAETORIAN score ≥90 with 98% (95% CI 97%-99%) negative predictive value. Conclusions: In contemporary practice, the PRAETORIAN score can be simplified. By adopting an intermuscular approach, two of the three steps of the score-evaluating the adequate posterior positioning of the generator and measuring the sub-generator fat-become superfluous, and impedance may serve as a reliable surrogate of sub-coil fat thickness. Furthermore, our data suggest that for non-obese patients, a favorable PRAETORIAN score is assured, making the score evaluation potentially unnecessary.
2025
conversion; defibrillation test; implantable defibrillator; subcutaneous
01 Pubblicazione su rivista::01a Articolo in rivista
The risk of failure of subcutaneous implantable cardioverter defibrillator therapy: from PRAETORIAN score to clinical practice / Ziacchi, Matteo; Ottaviano, Luca; Checchi, Luca; Viani, Stefano; Nigro, Gerardo; Bianchi, Valter; De Bonis, Silvana; De Filippo, Paolo; Francia, Pietro; Rapacciuolo, Antonio; Vitulano, Gennaro; Perego, Giovanni Battista; Schillaci, Vincenzo; Lavalle, Carlo; Migliore, Federico; Pisanò, Ennio C L; Compagnucci, Paolo; Palmisano, Pietro; Botto, Gianluca; Rordorf, Roberto; Lovecchio, Mariolina; Valsecchi, Sergio; Biffi, Mauro. - In: EUROPACE. - ISSN 1099-5129. - 27:2(2025). [10.1093/europace/euaf011]
File allegati a questo prodotto
File Dimensione Formato  
Ziacchi_Risk-failure_2025.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 3.53 MB
Formato Adobe PDF
3.53 MB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1732276
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 3
social impact