Aims Pulmonary vein isolation (PVI) for paroxysmal atrial fibrillation (PAF) using very high-power short-duration (vHPSD) radiofrequency (RF) ablation proved to be safe and effective. However, vHPSD applications result in shallower lesions that might not be always transmural. Multidetector computed tomography-derived left atrial wall thickness (LAWT) maps could enable a thickness-guided switching from vHPSD to the standard-power ablation mode. The aim of this randomized trial was to compare the safety, the efficacy, and the efficiency of a LAWT-guided vHPSD PVI approach with those of the CLOSE protocol for PAF ablation (NCT04298177). Methods and results Consecutive patients referred for first-time PAF ablation were randomized on a 1:1 basis. In the QDOT-by-LAWT arm, for LAWT <= 2.5 mm, vHPSD ablation was performed; for points with LAWT > 2.5 mm, standard-power RF ablation titrating ablation index (AI) according to the local LAWT was performed. In the CLOSE arm, LAWT information was not available to the operator; ablation was performed according to the CLOSE study settings: AI >= 400 at the posterior wall and >= 550 at the anterior wall. A total of 162 patients were included. In the QDOT-by-LAWT group, a significant reduction in procedure time (40 vs. 70 min; P < 0.001) and RF time (6.6 vs. 25.7 min; P < 0.001) was observed. No difference was observed between the groups regarding complication rate (P = 0.99) and first-pass isolation (P = 0.99). At 12-month follow-up, no significant differences occurred in atrial arrhythmia-free survival between groups (P = 0.88). Conclusion LAWT-guided PVI combining vHPSD and standard-power ablation is not inferior to the CLOSE protocol in terms of 1-year atrial arrhythmia-free survival and demonstrated a reduction in procedural and RF times.
Personalized pulmonary vein isolation with very high-power short-duration lesions guided by left atrial wall thickness: the QDOT-by-LAWT randomized trial / Falasconi, Giulio; Penela, Diego; Soto-Iglesias, David; Francia, Pietro; Saglietto, Andrea; Turturiello, Dario; Viveros, Daniel; Bellido, Aldo; Alderete, Jose; Zaraket, Fatima; Franco-Ocaña, Paula; Huguet, Marina; Cámara, Óscar; Vătășescu, Radu; Ortiz-Pérez, José-Tomás; Martí-Almor, Julio; Berruezo, Antonio. - In: EUROPACE. - ISSN 1532-2092. - 26:4(2024). [10.1093/europace/euae087]
Personalized pulmonary vein isolation with very high-power short-duration lesions guided by left atrial wall thickness: the QDOT-by-LAWT randomized trial
Francia, Pietro;
2024
Abstract
Aims Pulmonary vein isolation (PVI) for paroxysmal atrial fibrillation (PAF) using very high-power short-duration (vHPSD) radiofrequency (RF) ablation proved to be safe and effective. However, vHPSD applications result in shallower lesions that might not be always transmural. Multidetector computed tomography-derived left atrial wall thickness (LAWT) maps could enable a thickness-guided switching from vHPSD to the standard-power ablation mode. The aim of this randomized trial was to compare the safety, the efficacy, and the efficiency of a LAWT-guided vHPSD PVI approach with those of the CLOSE protocol for PAF ablation (NCT04298177). Methods and results Consecutive patients referred for first-time PAF ablation were randomized on a 1:1 basis. In the QDOT-by-LAWT arm, for LAWT <= 2.5 mm, vHPSD ablation was performed; for points with LAWT > 2.5 mm, standard-power RF ablation titrating ablation index (AI) according to the local LAWT was performed. In the CLOSE arm, LAWT information was not available to the operator; ablation was performed according to the CLOSE study settings: AI >= 400 at the posterior wall and >= 550 at the anterior wall. A total of 162 patients were included. In the QDOT-by-LAWT group, a significant reduction in procedure time (40 vs. 70 min; P < 0.001) and RF time (6.6 vs. 25.7 min; P < 0.001) was observed. No difference was observed between the groups regarding complication rate (P = 0.99) and first-pass isolation (P = 0.99). At 12-month follow-up, no significant differences occurred in atrial arrhythmia-free survival between groups (P = 0.88). Conclusion LAWT-guided PVI combining vHPSD and standard-power ablation is not inferior to the CLOSE protocol in terms of 1-year atrial arrhythmia-free survival and demonstrated a reduction in procedural and RF times.File | Dimensione | Formato | |
---|---|---|---|
Falasconi_Personalized-pulmonary_2024.pdf
accesso aperto
Tipologia:
Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza:
Creative commons
Dimensione
1.13 MB
Formato
Adobe PDF
|
1.13 MB | Adobe PDF |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.