Background: Robotic-assisted percutaneous coronary intervention (PCI) is a novel technology that permits remote operation of interventional devices. However, little is known about the safety and effectiveness of introducing a robotic PCI program in a hospital already experienced in traditional coronary angioplasty. Methods: Prospective single-arm survey to assess the safety and effectiveness of robotic-assisted PCI in comparison to pre-defined performance goals. The study cohort comprised all consecutive cases treated with robotic PCI since its introduction. The safety primary endpoint was a composite of (I) overall death or (II) non-fatal adverse events related to target vessel complications (stent thrombosis, myocardial infarction, vessel perforation or cardiac tamponade, or repeat invasive treatment) during the index hospitalization. The efficacy primary endpoint was robotic-assisted procedural success, a composite of (I) successful dilatation of the target lesion and (II) successful robotic assistance, defined as absent non-planned manual conversion. Results: A total of 83 patients and 112 lesions were prospectively enrolled. The rate of angiographic success was 99.1%. From these, 97 lesions (86.6%) were treated with only robotic PCI or with hybrid according to the pre-interventional plan. The rates of efficacy and safety primary endpoints were 85.7% and 2.4% respectively (P<0.01 for non-inferior to the pre-defined performance threshold). Conclusions: Introduction of robotic-assisted PCI in a tertiary center was associated with safe and effective results, comparable to pre-defined goals of optimal performance.

Safety and effectiveness of introducing a robotic-assisted percutaneous coronary intervention program in a tertiary center: a prospective study / Lemos, Pedro A.; Franken, Marcelo; Mariani Jr, Jose; Caixeta, Adriano; Almeida, Breno O.; Pitta, Fabio G.; Prado, Guy F. A.; Garzon, Stefano; Ramalho, Felipe; Albuquerque, Gabriel; Gomes, Ivanise M.; de Oliveira, Irisvaldo S.; Valle, Leonardo; Galastri, Leonardo; Affonso, Breno B.; Nasser, Felipe; Garcia, Rodrigo G.. - In: CARDIOVASCULAR DIAGNOSIS AND THERAPY. - ISSN 2223-3652. - 12:1(2022), pp. 67-76. [10.21037/cdt-21-442]

Safety and effectiveness of introducing a robotic-assisted percutaneous coronary intervention program in a tertiary center: a prospective study

Prado, Guy F. A.;
2022

Abstract

Background: Robotic-assisted percutaneous coronary intervention (PCI) is a novel technology that permits remote operation of interventional devices. However, little is known about the safety and effectiveness of introducing a robotic PCI program in a hospital already experienced in traditional coronary angioplasty. Methods: Prospective single-arm survey to assess the safety and effectiveness of robotic-assisted PCI in comparison to pre-defined performance goals. The study cohort comprised all consecutive cases treated with robotic PCI since its introduction. The safety primary endpoint was a composite of (I) overall death or (II) non-fatal adverse events related to target vessel complications (stent thrombosis, myocardial infarction, vessel perforation or cardiac tamponade, or repeat invasive treatment) during the index hospitalization. The efficacy primary endpoint was robotic-assisted procedural success, a composite of (I) successful dilatation of the target lesion and (II) successful robotic assistance, defined as absent non-planned manual conversion. Results: A total of 83 patients and 112 lesions were prospectively enrolled. The rate of angiographic success was 99.1%. From these, 97 lesions (86.6%) were treated with only robotic PCI or with hybrid according to the pre-interventional plan. The rates of efficacy and safety primary endpoints were 85.7% and 2.4% respectively (P<0.01 for non-inferior to the pre-defined performance threshold). Conclusions: Introduction of robotic-assisted PCI in a tertiary center was associated with safe and effective results, comparable to pre-defined goals of optimal performance.
2022
coronary; angioplasty; noninferiority; robotics; stent
01 Pubblicazione su rivista::01a Articolo in rivista
Safety and effectiveness of introducing a robotic-assisted percutaneous coronary intervention program in a tertiary center: a prospective study / Lemos, Pedro A.; Franken, Marcelo; Mariani Jr, Jose; Caixeta, Adriano; Almeida, Breno O.; Pitta, Fabio G.; Prado, Guy F. A.; Garzon, Stefano; Ramalho, Felipe; Albuquerque, Gabriel; Gomes, Ivanise M.; de Oliveira, Irisvaldo S.; Valle, Leonardo; Galastri, Leonardo; Affonso, Breno B.; Nasser, Felipe; Garcia, Rodrigo G.. - In: CARDIOVASCULAR DIAGNOSIS AND THERAPY. - ISSN 2223-3652. - 12:1(2022), pp. 67-76. [10.21037/cdt-21-442]
File allegati a questo prodotto
File Dimensione Formato  
Lemos_Safety-effectiveness-introducing_2022.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 688.01 kB
Formato Adobe PDF
688.01 kB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1716296
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact