Purpose: To assess the impact of different quantum iterative reconstruction (QIR) levels on objective and subjective image quality of ultra -high resolution (UHR) coronary CT angiography (CCTA) images and to determine the effect of strength levels on stenosis quantification using photon -counting detector (PCD)-CT. Method: A dynamic vessel phantom containing two calcified lesions (25 % and 50 % stenosis) was scanned at heart rates of 60, 80 and 100 beats per minute with a PCD-CT system. In vivo CCTA examinations were performed in 102 patients. All scans were acquired in UHR mode (slice thickness0.2 mm) and reconstructed with four different QIR levels (1 - 4) using a sharp vascular kernel (Bv64). Image noise, signal-to-noise ratio (SNR), sharpness, and percent diameter stenosis (PDS) were quantified in the phantom, while noise, SNR, contrast -tonoise ratio (CNR), sharpness, and subjective quality metrics (noise, sharpness, overall image quality) were assessed in patient scans. Results: Increasing QIR levels resulted in significantly lower objective image noise (in vitro and in vivo: both p < 0.001), higher SNR (both p < 0.001) and CNR (both p < 0.001). Sharpness and PDS values did not differ significantly among QIRs (all pairwise p > 0.008). Subjective noise of in vivo images significantly decreased with increasing QIR levels, resulting in significantly higher image quality scores at increasing QIR levels (all pairwise p < 0.001). Qualitative sharpness, on the other hand, did not differ across different levels of QIR (p = 0.15). Conclusions: The QIR algorithm may enhance the image quality of CCTA datasets without compromising image sharpness or accurate stenosis measurements, with the most prominent benefits at the highest strength level.

Ultra-high resolution coronary CT angiography on photon-counting detector CT: bi-centre study on the impact of quantum iterative reconstruction on image quality and accuracy of stenosis measurements / Vecsey-Nagy, M.; Varga-Szemes, A.; Schoepf, U. J.; Tremamunno, G.; Fink, N.; Zsarnoczay, E.; Szilveszter, B.; Graafen, D.; Halfmann, M. C.; Vattay, B.; Boussoussou, M.; O'Doherty, J.; Suranyi, P. S.; Maurovich-Horvat, P.; Emrich, T.. - In: EUROPEAN JOURNAL OF RADIOLOGY. - ISSN 1872-7727. - 176:(2024). [10.1016/j.ejrad.2024.111517]

Ultra-high resolution coronary CT angiography on photon-counting detector CT: bi-centre study on the impact of quantum iterative reconstruction on image quality and accuracy of stenosis measurements

Tremamunno G.;
2024

Abstract

Purpose: To assess the impact of different quantum iterative reconstruction (QIR) levels on objective and subjective image quality of ultra -high resolution (UHR) coronary CT angiography (CCTA) images and to determine the effect of strength levels on stenosis quantification using photon -counting detector (PCD)-CT. Method: A dynamic vessel phantom containing two calcified lesions (25 % and 50 % stenosis) was scanned at heart rates of 60, 80 and 100 beats per minute with a PCD-CT system. In vivo CCTA examinations were performed in 102 patients. All scans were acquired in UHR mode (slice thickness0.2 mm) and reconstructed with four different QIR levels (1 - 4) using a sharp vascular kernel (Bv64). Image noise, signal-to-noise ratio (SNR), sharpness, and percent diameter stenosis (PDS) were quantified in the phantom, while noise, SNR, contrast -tonoise ratio (CNR), sharpness, and subjective quality metrics (noise, sharpness, overall image quality) were assessed in patient scans. Results: Increasing QIR levels resulted in significantly lower objective image noise (in vitro and in vivo: both p < 0.001), higher SNR (both p < 0.001) and CNR (both p < 0.001). Sharpness and PDS values did not differ significantly among QIRs (all pairwise p > 0.008). Subjective noise of in vivo images significantly decreased with increasing QIR levels, resulting in significantly higher image quality scores at increasing QIR levels (all pairwise p < 0.001). Qualitative sharpness, on the other hand, did not differ across different levels of QIR (p = 0.15). Conclusions: The QIR algorithm may enhance the image quality of CCTA datasets without compromising image sharpness or accurate stenosis measurements, with the most prominent benefits at the highest strength level.
2024
Coronary computed tomography angiography; Photon-counting detector; Quantum iterative reconstruction; Ultra-high resolution
01 Pubblicazione su rivista::01a Articolo in rivista
Ultra-high resolution coronary CT angiography on photon-counting detector CT: bi-centre study on the impact of quantum iterative reconstruction on image quality and accuracy of stenosis measurements / Vecsey-Nagy, M.; Varga-Szemes, A.; Schoepf, U. J.; Tremamunno, G.; Fink, N.; Zsarnoczay, E.; Szilveszter, B.; Graafen, D.; Halfmann, M. C.; Vattay, B.; Boussoussou, M.; O'Doherty, J.; Suranyi, P. S.; Maurovich-Horvat, P.; Emrich, T.. - In: EUROPEAN JOURNAL OF RADIOLOGY. - ISSN 1872-7727. - 176:(2024). [10.1016/j.ejrad.2024.111517]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1713333
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact