Objectives: Calibration is an important source of variability in liquid chromatography mass spectrometry (LC-MS) methods for insulin-like growth factor 1 (IGF-1). This study investigated the impact of different calibrator matrices on IGF-1 measurements by LC-MS. Moreover, the comparability of immunoassays and LC-MS was assessed.Design & Methods: Calibrators from 12.5 to 2009 ng/ml were prepared by spiking WHO international Standard (ID 02/254 NIBSC, UK) into the following matrices: native human plasma, fresh charcoal-treated human plasma (FCTHP), old charcoal-treated human plasma, deionized water, bovine serum albumin (BSA), and rat plasma (RP). A validated in-house LC-MS method was calibrated repeatedly with these calibrators. Then, serum samples from 197 growth hormone excess and deficiency patients were analysed with each calibration.Results: The seven calibration curves had different slopes leading to markedly different patient results. The largest differences in IGF-1 concentration from the median (interquartile range) was observed with the calibrator in water and the calibrator in RP (336.4 [279.6-417.0] vs. 112.5 [71.2-171.2], p < 0.001). The smallest difference was observed with calibrators in FCTHP and BSA (141.8 [102.0-198.5] vs. 127.9 [86.9-186.0], p < 0.049). Compared to LC-MS with calibrators in FCTHP, immunoassays showed relevant proportional bias (range:-43% to-68%), constant bias (range: 22.84 to 57.29 ng/ml) and pronounced scatter. Comparing the immunoassays with each other revealed proportional bias of up to 24%.Conclusions: The calibrator matrix is critical for the measurement of IGF-1 by LC-MS. Regardless of the calibrator matrix, LC-MS shows poor agreement with immunoassays. Also, the agreement between different immunoassays is variable.

The impact of different calibration matrices on the determination of insulin-like growth factor 1 by high-resolution-LC-MS in acromegalic and growth hormone deficient patients / Simstich, Sebastian; Züllig, Thomas; D'Aurizio, Federica; Biasotto, Alessia; Colao, Annamaria; Isidori, Andrea M; Lenzi, Andrea; Fauler, Günter; Köfeler, Harald C; Curcio, Francesco; Herrmann, Markus. - In: CLINICAL BIOCHEMISTRY. - ISSN 1873-2933. - 114:(2023). [10.1016/j.clinbiochem.2023.02.008]

The impact of different calibration matrices on the determination of insulin-like growth factor 1 by high-resolution-LC-MS in acromegalic and growth hormone deficient patients

Isidori, Andrea M;Lenzi, Andrea;
2023

Abstract

Objectives: Calibration is an important source of variability in liquid chromatography mass spectrometry (LC-MS) methods for insulin-like growth factor 1 (IGF-1). This study investigated the impact of different calibrator matrices on IGF-1 measurements by LC-MS. Moreover, the comparability of immunoassays and LC-MS was assessed.Design & Methods: Calibrators from 12.5 to 2009 ng/ml were prepared by spiking WHO international Standard (ID 02/254 NIBSC, UK) into the following matrices: native human plasma, fresh charcoal-treated human plasma (FCTHP), old charcoal-treated human plasma, deionized water, bovine serum albumin (BSA), and rat plasma (RP). A validated in-house LC-MS method was calibrated repeatedly with these calibrators. Then, serum samples from 197 growth hormone excess and deficiency patients were analysed with each calibration.Results: The seven calibration curves had different slopes leading to markedly different patient results. The largest differences in IGF-1 concentration from the median (interquartile range) was observed with the calibrator in water and the calibrator in RP (336.4 [279.6-417.0] vs. 112.5 [71.2-171.2], p < 0.001). The smallest difference was observed with calibrators in FCTHP and BSA (141.8 [102.0-198.5] vs. 127.9 [86.9-186.0], p < 0.049). Compared to LC-MS with calibrators in FCTHP, immunoassays showed relevant proportional bias (range:-43% to-68%), constant bias (range: 22.84 to 57.29 ng/ml) and pronounced scatter. Comparing the immunoassays with each other revealed proportional bias of up to 24%.Conclusions: The calibrator matrix is critical for the measurement of IGF-1 by LC-MS. Regardless of the calibrator matrix, LC-MS shows poor agreement with immunoassays. Also, the agreement between different immunoassays is variable.
2023
Hormone; Insulin-like Growth Factor 1 (IGF-1); Liquid Chromatography High-Resolution Mass Spectrometry (LC-HRMS); Online Solid Phase Extraction (OSPE); Peptide; Quantification; Serum
01 Pubblicazione su rivista::01a Articolo in rivista
The impact of different calibration matrices on the determination of insulin-like growth factor 1 by high-resolution-LC-MS in acromegalic and growth hormone deficient patients / Simstich, Sebastian; Züllig, Thomas; D'Aurizio, Federica; Biasotto, Alessia; Colao, Annamaria; Isidori, Andrea M; Lenzi, Andrea; Fauler, Günter; Köfeler, Harald C; Curcio, Francesco; Herrmann, Markus. - In: CLINICAL BIOCHEMISTRY. - ISSN 1873-2933. - 114:(2023). [10.1016/j.clinbiochem.2023.02.008]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1706894
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 2
social impact